beautypg.com

Rear panel usb port interface, Age 93) – GW Instek GDS-1000-U Series User Manual User Manual

Page 93

background image

CONFIGURATION

93

Rear Panel USB Port Interface

The Remote control interface section describes how to set up the
USB interface for PC connection. The details of remote control
commands are described in the GDS-1000-U Programming Manual.

USB connection PC end

Type A, host

GDS-1000-U end

Type B, slave

Speed

1.1/2.0 (full speed)

Procedure

1. Connect the USB cable to

the USB slave port on the
GDS-1000-U.

2. When the PC asks for the USB driver, select

dso_cdc_1000.inf which is downloadable from
the GW website, www.gwinstek.com.tw, GDS-
1000-U product corner.

3. On the PC, activate a terminal application such

as MTTTY (Multi-Threaded TTY). To check the
COM port No., see the Device Manager in the
PC. For WindowsXP, select Control panel →
System → Hardware tab.

4. Run this query command via the terminal

application.
*idn?
This command should return the manufacturer,
model number, serial number, and firmware
version in the following format.
GW, GDS-1052-U, 000000001, V1.00

5. Configuring the command interface is

completed. Refer to the programming manual
for the remote commands and other details.