7. The Channel1 calibration

will complete in less than 5
minutes.

Ch1 calibration 1/3

8. When finished, connect the

calibration signal to the
Channel 2 input and repeat
the procedure.

Done!!

9. When the calibration is complete the display

will go back to the previous state.

Probe Compensation

Procedure

1. Connect the probe between the Channel1 input

and the probe compensation output (2Vp-p,
1kHz square wave) on the front panel. Set the
probe voltage attenuation to x10.

VOLTS/DIV

VOLTS/DIV

TIME/DIV

CH 1

MATH

CH 2

MENU

MENU

Acquire

Display

Utility

Help

Run/Stop

VARIABLE

FORCE

Autoset

Cursor

SINGLE

Hardcopy

Measure

Save/Recall

LEVEL

VERTICAL

HORIZONTAL

TRIGGER

CH1

CAT

300V

M

W

15pF

MAX. 300Vpk

1

CH2

EXT TRIG

CAT

300V

M

W

15pF

MAX. 300Vpk

1

X

Y

X

1

0

X

1

x1

x10

2. Press the Utility key.

Utility

3. Press ProbeComp.

ProbeComp

Menu