Selecting the waveform update mode – GW Instek GDS-1000-U Series User Manual User Manual

Page 79

background image

CONFIGURATION

79

Selecting the waveform update mode

Background

The display update mode is switched
automatically or manually according to the
horizontal scale.

Main mode

Updates the whole displayed waveform at once.
The main mode is automatically selected when the
horizontal scale (timebase) is fast.

Horizontal scale

≤100ms/div

Trigger

All modes available

Roll mode

Updates and moves the waveform gradually from
the right side of the display to the left. The Roll
mode is automatically selected when the
horizontal scale (timebase).
When in the Roll mode, an indicator appears at the
bottom of the display.

Main mode

Roll mode

Timebase

≥50ms/div (≤5kS/s)

Trigger

Auto mode only

Selecting the Roll
mode manually

1. Press the Horizontal menu

key.

MENU

2. Press Roll. The horizontal

scale automatically becomes
50ms/div and the
waveform starts scrolling
from the right side of the
display (If the oscilloscope
is already in the Roll mode,
there will be no change).

Roll