<6. OPERATION OF SC (Conductivity)>

6-7

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

n

Contact status

This screen is displayed only if the PH201G distributor is used and “PH201G” is selected in

communication setup.

n

c.c. (factory)

This parameter displays the cell constant (factory setting).
The cell constant (factory setting) is determined by factory calibration made during sensor

manufacturing. This value can be set in Commissioning → Measurement setup → Sensor setup.
The cell constant is indicated on the sensor or cable label.

n

c.c. (adjusted)

This parameter displays the cell constant (adjusted).
The cell constant (adjusted) is a cell constant that has been calibrated and is set by calibration

action.
When the system’s cell constant is adjusted on-line by calibration using a process solution or

buffer solution, a new cell constant is logged here. The difference between this value and the

default value set at factory shipment will not be large. If there is a large difference between them,

check whether the sensor is broken or contaminated.

n

Temp. comp. 1

This parameter shows the temperature compensation method for 1st measurement. The setting

is made in Commissioning → Measurement setup → Temp. comp.

n

Temp. comp. 2

This parameter shows the temperature compensation method for 2nd measurement. The setting

is made in Commissioning → Measurement setup → Temp. comp.
2nd measurement does not imply that two measurements can be made individually, but it means

that two types of compensation methods can be configured, which enables two phases to be

monitored accurately in the same process. An example is monitoring the switching between

process and washing solutions.

n

Polarization

This parameter indicates the degree of polarization of the sensor. It is an index of the degree

of sensor contamination (adherents). The greater the value, the more the polarization has

progressed. If the value is high, maintenance should be performed on the sensor as soon as

possible.

n

Sensor Ohm

This parameter represents the non-compensated resistance of the sensor.

n

USP

USP stands for United States Pharmacopeia and indicates the margin of safety for the water for

injection (WFI) conductivity at each temperature as a percentage.