4 regulatory compliance, Regulatory compliance -9, 1ph sc isc do – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual
Page 21: Sencom
<1. INTRODUCTION AND GENERAL DESCRIPTION>
1-9
IM 12A01A02-01E
5th Edition : Oct. 31, 2013-00
1.4
Regulatory Compliance
Safety: EN61010-1
UL 61010-1
CAN/CSA C22.2 No.61010-1
EMC: EN61326-1 Class A, Table 2 (For use in industrial locations)
EN61326-2-3
AS/NZS CISPR11
Korea Electromagnetic Conformity Standard
Installation altitude:
2000 m or less
Category based on IEC 61010: I (Note 1)
Pollution degree based on IEC 61010: 2 (Note 2)
Note 1: Installation category, called over-voltage category, specifies impulse withstand voltage.
Equipment with “Category I” (ex. two-wire transmitter) is used for connection to circuits in which measures are taken
to limit transient over-voltages to an appropriately low level.
Note 2: Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce
dielectric strength. Degree 2 is the normal indoor environment.
Explosion-proof (Intrinsically safe type and non-incendive) (for suffix code: -EA):
ATEX Intrinsically safe approval
Applicable standard
Explosive Atmospheres
EN 60079-0:2009
General requirements
EN 60079-11:2007
Equipment protection by intrinsic safety “i”
EN 60079-26:2007
Equipment with equipment protection level (EPL) Ga
EN 60529:1992
Degrees of protectionprovided by enclosures (IP Code)
Type of protection
II 1G Ex ia IIC Ga
Group:
II
Category:
1G
T4: for ambient temperature:–20 to 55ºC
T6: for ambient temperature:–20 to 40ºC
Atmosphere pressure: 80 kPa (0.8 bar) to 110 kPa (1.1 bar)
Degree of Protection of the Enclosure: IP66
IECEx Intrinsically safe
Applicable standard
IEC 60079-0:2007
Part 0: General requirements
IEC 60079-11: 2006 Part 11: Equipment protection by intrinsic safety “i”
IEC 60079-26: 2006 Part 26: Construction, test and marking of Group II Zone 0
electrical
apparatus
IEC 60529: 2001
Degrees of protection provided by enclosures (IP Code)
Type of protection
Ex ia IIC Ga
T4: for ambient temperature:–20 to 55ºC
T6: for ambient temperature:–20 to 40ºC
Atmosphere pressure: 80 kPa (0.8 bar) to 110 kPa (1.1 bar)
Degree of Protection of the Enclosure: IP66
FM Intrinsically safe and nonincentive approval
Applicable standard
FM-3600: 2011 Approval Standard for Electric Equipment for use in Hazardous
(Classified) Locations General Requirement
FM-3610: 2010 Approval Standard for Intrinsically Safe Apparatus and Associated
Apparatus for Use in Class I, II, and III, Division 1, Hazardous
(Classified) Locations
FM-3611: 2004 Nonincendive Electrical Equipment for Use in Class I and II, Division 2
and Class III, Divisions 1 and 2, Hazardous (Classified) Locations
FM-3810: 2005 Electrical Equipment for Measurement, Control and Laboratory Use
NEMA 250:1991 Enclosures for Electrical Equipment (1000 Volts Maximum)
ANSI/IEC 60529:2004
Degrees of protection provided by enclosures (IP Code)
ANSI/ISA 60079-0 2009
Part 0: General Requirements
ANSI/ISA 60079-11 2011
Part 11: Equipment protection by intrinsic safety “i”
1
PH
SC
ISC
DO
SENCOM