<11. CALIBRATION OF ISC (Inductive Conductivity)>

11-2

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

NOTE

The standard instrument to be used in calibration with a process solution should always be

accurate. Yokogawa recommends that the Model SC72 pocket conductivity meter be used for

this purpose.

Where temperature compensation 1 (SC1) and temperature compensation 2 (SC2) have been

configured, the configured temperature compensation is effective even during calibration.

Therefore, the reading is the value converted to a conductivity value at the reference temperature

set in Temperature settings.
There are temperature compensation 1 (SC1) and temperature compensation 2 (SC2), but this

does not mean that calibration is required twice. It means that either SC1 or SC2 temperature

compensation should be selected and calibration should be made once to obtain the cell

constant. The cell constant after calibration can be checked on the Detail screen.

NOTE

When a sensor is exchanged or replaced, sensor wellness data should be reset.
When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to

the figure 9.8.)

l

Cell constant

The center value of the cell constant of the ISC40 sensor is 1.88 cm

-1

. The nominal cell constants

of individual sensors are indicated on the cable markers and the actual installation can change

this factor. If there is less than 30 mm spacing between a sensor and holder, in-situ calibration is

necessary to meet the specified accuracies.

• If the sensor is installed in the stainless steel standard holder ISC40FFJ-S, the cell constant

is reduced by approximately 7%; enter a value 7% smaller than the value on the marker of

the sensor cable.

• If the sensor is installed in the polypropylene standard holder ISC40FFJ-P, the cell constant

is increased by approximately 1%; enter a value 1% greater than the value on the sensor

cable marker.

• If the sensor is installed in piping that is long in the axial direction with the cross section

as shown in Figure 11.3, the cell constant of the sensor installed in the piping (reference

data for a design center value of 1.88 cm

-1

) is as shown in Figure 11.3. A value obtained by

multiplying the value on the sensor cable marker and a value read from Figure 11.3 should

be entered.

X= 30 mm (Min.)

X

X

X

Figure 11.2

Sensor in calibration solution