6 factory setup, 6 display setup, 1 main display – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual
Page 134: 2 trend, Factory setup -11, Display setup -11 10.6.1, Main display -11, 2 trend -11, Nph201g, Isc 10
<10. COMMISSIONING OF ISC (Inductive Conductivity)>
10-11
IM 12A01A02-01E
5th Edition : Oct. 31, 2013-00
n
PH201G
Select this menu if the PH201G distributor is connected to the instrument.
In the PH201G setup screen, make settings for “Hold contact,” and “Fail contact.”
l
Hold contact
Select Disabled or Enabled.
When this item is enabled, the output will be held according to the setting of “Hold type” on the
Hold setup screen.
l
Fail contact
Select a status from among “Fail + Warn,” “Fail only,” and “Disabled.”
This setting depends on the error configuration. See “10.3 Error configuration.”
“Fail” corresponds to “Fault.”
10.5.6
Factory setup
For “Factory setup,” there is no item to be set by the user.
NOTE
This menu is for service engineers only. This section is protected by a password. Attempting to
change data in the factory adjustment menu without the proper instructions and equipment could
result in corruption of the instrument setup and damage the performance of the unit.
10.6 Display setup
This screen is used to make various settings relating to screen display.
NOTE
Settable items differ depending on settings in “Measurement setup.”
10.6.1
Main display
Three measurement values can be set to display on the Main display as a primary value (1st
line), a second value (2nd line) and a third value (3rd line) respectively.
On the “Additional text”, a text of up to 12 alphanumeric characters can be assigned to each
measurement value.
Additional texts are displayed on the Main display, and are useful for identifying measurements.
In some cases, not all 12 characters can be displayed due to the letters; check the texts
displayed on the Main display after setting. If a part of the text is missing, adjust the number of
characters.
10.6.2 Trend
This screen is used to make settings for the Trend Graph Screen.
Set the process parameters to be displayed for each trend. They can be set for the 1st to 3rd
trends. When all three process parameters are set “Empty”, there is no trend display (no trend
button).
ISC