<2. WIRING AND INSTALLATION>

2-15

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

2.7 Operation

When all wiring is completed, turn on the power of the instrument. Make sure that the LCD

screen turns on. Then, the screen changes to the Quick Setup display. If the instrument is not

configured, an error indicator may be displayed, or the measurement values displayed may be

incorrect. Check the initial settings and change them to suit your purpose.

If you are using the without display unit, refer to “HART communication (TI 12A01A02-60E).”

NOTE

After turning on the power, it takes occasionally more than 1 minute to change to the Quick Setup

display.

NOTE

A current of 12 mA is output for several seconds after power-on. This is the specification.

l

Change language

When the power is turned on, the Quick setup display appears.
Upon delivery, the display language is set to English. If you want to use a language other than

English, select your desired language.

Chinese
Czech
English
French
German
Italian
Japanese
Korean
Portuguese
Russian
Spanish

Change language

The instrument will restart
Are you sure?

Yes

Warning

Start quick setup?

Yes
No

Change language

Quick setup

Chinese
Czech
English
French
German
Italian
Japanese
Korean
Polish
Portuguese
Russian
Spanish

(2)

(3)

(1)

Figure 2.10

Changing the language

(1) Select ♦Change language.
(2) Select a desired language (click the

to scroll the language options.)

(3) Click YES.
Changing languages automatically reboots the FLXA21.

Check that the parameters match your specifications and modify them if necessary before using

the instrument.
A sheet for user settings is contained in the CD-ROM. Please take a memo of your settings on it

when you configure the FLXA21 and keep the sheet just in case.