<15. OPERATION OF SENCOM pH/ORP>

15-7

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

n

PH (ORP)’s zero, slope, and sensor, and Impedance

l

Zero

= calibrated sensor offset in mV. Theoretically, the sensor reads 0 mV in a buffer solution of pH

7. The ZERO value indicates the condition of the sensor. The trend of ZERO drift of the sensor is

used to predict the lifetime of the sensor.
ZERO can also be displayed in pH units and then it represents the pH value where the sensor

output is 0 mV at 25ºC. Setting can be made in Commissioning → Measurement setup →

Calibration settings → Zero and Slope units.
Setting of the zero value can be made in Commissioning → Measurement setup → Calibration

settings → Zero/Slope/ITP. For details, see section 16.2.4.

l

Slope

= calibrated efficiency of the sensor unit as a percentage of the theoretical slope of the sensor

unit. The theoretical slope follows the NERNST equation and is 59.16 mV/pH (at 25ºC). The

SLOPE can be calibrated only after a two-point calibration in buffer solutions with a different pH

value. A low slope indicates that the sensor is not clean or is faulty.
The SLOPE can also be displayed as a mV/pH value at 25ºC if the user has defined this variable

as mV/pH in Commissioning → Measurement setup → Calibration settings → Zero and Slope

units.
Setting of the slope value can be made in Commissioning → Measurement setup → Calibration

settings → Zero/Slope/ITP. For details, see section 16.2.4.

pH14

pH0

pH7, 0mV

Theoretical value

+mV

-mV Calibration curve (slope)

Zero

F0410.ai

Figure 15.6

Zero/slope

l

Sensor

Represents the electromotive force of the sensor.

l

Impedance 1

For a PH sensor, “Impedance” shows electrical resistance of the glass membrane electrode. The

FLXA21 checks the impedance to know damage of the electrode.
For an ORP sensor,”Impedance” shows electrical resistance of metal electrode.
The FLXA21 checks the impedance to know the surface condition of smudge and the snapping

of sensor wires. In case of “Input Impedance setting” is “High” and the measured input-1

impedance value is higher than 100 kΩ, the display shows “MΩ RANGE”. The measured input-1

impedance value is lower than 100 kΩ, display shows “BAD”.
If both impedance measurements are disabled (Error setting: Off), the display shows “- - - - (bar)”.

If either impedance measurement 1 or 2 is enabled, the display shows both the impedance

values.