Norp settings, Ph 4 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 61

background image

<4. COMMISSIONING OF pH/ORP>

4-7

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

l

Auto correct (Zero, Slope)

This function calculates calibration coefficients from the transition of past calibration data (zero,

slope) and compensates pH value using these calibration coefficients after the latest calibration.

The default setting is “Disable.” To make the function effective, select “Enable.”

Time

Calibration

Zero, Slope

a

b

c

d

a: Actual change

b: Zero and slope calculated from

the calibration data

c: Zero and slope calculated from

the calibration data and shifted

using the latest calibration data

d: Data from calibration

Figure 4.3

Auto correct

n

ORP settings

l

Limits and timing

Zero High/Low
Zero (aspot) high and low limits. During calibration, it is checked whether the new zero exceeds

these high and low limits. Narrowing the band will prevent bad calibration procedures and

calibration of bad sensors, resulting in higher accuracy. The default values should be adjusted to

suit the application and the “users” criterion.

Slope High/Low
Slope (sensitivity) high and low limits. During calibration, it is checked whether the new slope

exceeds these high and low limits. Narrowing the band will prevent bad calibration procedures

and calibration of bad sensors, resulting in higher accuracy. The default values should be

adjusted to suit the application and the “users” criterion.

Step Range
Set the range over which the stability of a measured value is checked. If variations of a measured

value over the stabilization time are within this setpoint, the measured value is judged to have

stabilized.

Stabilization time
During calibration, the stability of the ORP value is constantly monitored. When variations of the

ORP value are within a value set in Step Range over this stabilization time set here, the value

is regarded as being stable. If the ORP value does not stabilize within 10 minutes, calibration is

aborted.

Calibr. interval
Set the interval in which a new calibration must take place. If the interval set here is exceeded,

the instrument will be notified according to the setting in “Calib. time exceeded” in the error

configuration.

l

Zero/Slope

You can enter zero and slope values directly.

PH

4