<1. INTRODUCTION AND GENERAL DESCRIPTION>

1-8

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

For some errors concerned with setting, a message on the remedy can jump to the relevant

setting directly.

Problem with mA and Process
parameter.

Remedy

Select a correct ‘Process
parameter’ for mA in:
Commissioning >> Output
setup

Fault:
mA configuration error

Figure 1.10

Example of jumping from remedy to setting

If neither a fault is detected nor warning is indicated, “Working properly” is displayed.
If a fault is detected or warning is indicated, up to 3 messages appear.

n

Execute & Setup

Allows you to calibrate and configure the instrument. These operations can be protected with a

password.

Execute:
Calibration
HOLD
Temporary output

Setup:
Commissioning
Change language
Start Quick Setup

Execute & Setup

Figure 1.11

Example of Execute & Setup display

1.3 Passwords

Passwords can protect the FLXA21 from other persons’ access except a right operator’s.

Execute:
Calibration/Wash
HOLD
Temporary output

Setup:
Comissioning
Change language
Start Quick Setup

Execute &Setup

Sensor setup
Measurement setup
Output setup
Error configuration
Logbook configuration
Advanced setup
Display setup
Calculated data setup

Commissioning

Settings
Tag
Passwords
Date/Time
Communication
Factory setup

Advanced setup

Execute:

Commissioning:

Entering an empty password
results in disabling the
password check.

Passwords

Figure 1.12

Example of passwords setting

Execute operation and Commissioning operation can be separately protected by each password.

When the Execute password is set, execution of Calibration(/Wash), HOLD or Temporary output

needs to enter the password to start each execution. When the Commissioning password is set,

Commissioning and Start Quick Setup needs to enter the password to start the setting-up.

Refer to the section 4.6.3, 7.5.3, 10.5.3, 13.6.3 or 16.6.3 for details of setting passwords.