Nwiring of hamilton sensors, Ph sc isc do, Sencom – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual
Page 36
<2. WIRING AND INSTALLATION>
2-9
IM 12A01A02-01E
5th Edition : Oct. 31, 2013-00
n
Wiring of HAMILTON sensors
Sensor
Measure-
ment
Board Terminals
Input 2 Input 1 Jumper Remark
11
12
13
14
15
16
17
18 18/13 19/15
CHEMTRODE
pH
Blue
Red
(White)
Yes
No 13/14 WU20D
Cable
CHEMTRODE-
ORP
ORP
Yes
Yes 13/14
CHEMTRODE-
VP
pH
E
F
B
A
Yes
No 13/14 VP6-SC
CLARITRODE
pH
Blue
Red
(White)
Yes
No 13/14 WU20D
Cable
CLARITRODE-
VP
pH
E
F
B
A
Yes
No
VP6-SC
CONDCUELL
SC
White/
Grey Blue Green Pink Brown Yellow
Fixed
Cable
CONDCUELL-
VP
SC
Blue Yellow Black Black
Shield Grey
Grey
Shield
Hamilton
VP Cable
EASYFERM
pH
Blue
Red
(White)
Yes
No 13/14 WU20D
Cable
EASYFERM-
VP
pH
E
F
B
A
Yes
No 13/14 VP6-SC
INCHTRODE
pH
E
F
B
D
A
Yes
No
VP6-SC
MECOTRODE
pH
Blue
Red
(White)
Yes
No 13/14 WU20D
Cable
MECOTRODE-
VP
pH
E
F
B
A
Yes
No
VP8-DC
OXYFERM-
VP/
OXYGOLD
DO
White Green
Green/
Yellow
Black Red
Hamilton
VP Cable
OXYSENS
DO
Yellow Blue
Black
Clear Brown
Fixed
Cable
pHEASY
pH
E
F
B
D
A
B
Yes
No
POLILYTE
pH
Blue
Red
(White)
Yes
No
WU20D
Cable
POLILYTE-
VP
pH
White Green Red Blue Black/
Clear
Yes
No
Hamilton
VP Cable
POLILYTE
PLUS-VP
pH
White Green Red Blue Black/
Clear
Yes
No
Hamilton
VP Cable
Color coding of Variopin cables
Cable
PIN
A
B
C
D
E
F
G
H
Hamilton
VP6-SC
Black/
Clear
Red
Grey
Blue
White
Green
Hamilton
VP8-DC
Black/
Clear
Black
Shield
Red/
Clear
Red
Shield
White
Green
Yellow
Brown
WU20-V-S
VP6-SC
Clear
Brown
Black
Yellow
Red
Blue
WU20-V-D
VP8-DC
Brown
Core
Brown
Shield
White
Cored
White
Shield
Red
Blue
Yellow
PH
SC
ISC
DO
SENCOM
2