App.2-11, App. sc – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual
Page 227
App.2-11
IM 12A01A02-01E
5th Edition : Oct. 31, 2013-00
FLXA21 is programmed with the following table of conductivity of Potassium Chloride (KCl)
solutions at 25°C. This is used in the Automatic Cell Constant setting calibration feature. (See
chapter 9 on calibration) The table is derived from the Standards laid down in “International
Recommendation No. 56 of the Organisation Internationale de Métrologie Legale”.
Table 4 KCl values at 25 °C
Standard solution
mol/l
mg KCl/kg of solution
Conductivity
1.000 M KCl
1.0
71135.2
111.31 mS/cm
0.100 M KCl
0.1
7419.13
12.852 mS/cm
0.010 M KCl
0.01
745.263
1.4083 mS/cm
0.005 M KCl
0.005
373.29
0.7182 mS/cm
0.002 M KCl
0.002
149.32
0.2916 mS/cm
0.001 M KCl
0.001
74.66
0.1469 mS/cm
If it is more convenient, the user may make solutions from Sodium Chloride (NaCl or common
table salt) with the help of the following relationship table. This table is derived from the IEC norm
60746-3.
Table 5 NaCl values at 25 °C
Weight %
mg/kg
Conductivity
0.001
10
21.4 µS/cm
0.003
30
64.0 µS/cm
0.005
50
106 µS/cm
0.01
100
210 µS/cm
0.03
300
617 µS/cm
0.05
500
1.03 mS/cm
0.1
1000
1.99 mS/cm
0.3
3000
5.69 mS/cm
0.5
5000
9.48 mS/cm
1
10000
17.6 mS/cm
3
30000
48.6 mS/cm
5
50000
81.0 mS/cm
10
100000
140 mS/cm
For resistivity measurement the standard resistivity units of the calibration solution can be
calculated as follows:
R = 1000/G (kΩ•cm if G = µS/cm)
Example: 0.001% weight
R = 1000/21.4 = 46.7 kΩ•cm
n
Monitoring of Purified Water and WFI according to USP<645>
First published in the USP23 (The United States Pharmacopeia, ed. 23), the USP <645>
describes a procedure of measuring the conductivity of Purified Water and WFI (Water for
Injection) instead of the previous complicated monitoring procedure. This procedure consists
of three stages for measuring the conductivity of Purified Water and WFI. On the stage 1,
firstly the temperature of the water and the non-temperature-compensated conductivity of the
water are measured, and this conductivity is compared with the limit conductivity value of the
corresponding temperature directed by USP<645>. (Refer to Figure 3.) If the conductivity is
higher than the limit value, the procedure will proceed to the stage 2.
The FLXA21 has the limit values of USP<645> built in the firmware. When the error configuration
is set for this monitoring, the FLXA21 checks the non-temperature-compensated conductivity
with the limit value. If the conductivity is higher than the limit value, an error will be generated. A
safety margin to the limit value can be set on the FLXA21 to generate an error below limit value.
(Refer to the section 7.3.)
When the safety margin is set at 20%, for example, an error will be generated when the non-
temperature-compensated conductivity goes higher than the 80% of the limit value at all
temperatures. For example, if the temperature is 64 ºC. and the safety margin is set at 20%,
then an error will be generated at 0.8 x 2.2 μS/cm. = 1.76 μS/cm. (2.2 μS/cm is the USP<645>
limit value at 64ºC). In resistivity mode, an error will be generated at an non-temperature-
compensated resistivity of 0.568 MΩ (=1/1.76 μS/cm).
App.
SC