beautypg.com

4 sample, 5 temperature coefficient, 6 temperature calibration – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 111: 4 sample -3 8.5, Temperature coefficient -3, Temperature calibration -3, Sc 8

background image

<8. CALIBRATION OF SC (Conductivity)>

8-3

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

8.4 Sample

With the sensor in situ, a sample can be taken for laboratory analysis. Sample calibration records

the time and reading, and holds these in memory until the analysis has been completed. The

laboratory data can then be entered regardless of the current process value, without the need for

calculations.
When sample calibration is made with temperature compensation activated, the types of

temperature compensation for laboratory analysis equipment should be matched. Use of a

different type of temperature compensation between equipment causes an error. The standard

conductivity meter to be used should always be based on the accurate and same temperature

compensation calculation method. Yokogawa recommends that the Model SC82 pocket

conductivity meter be used for this purpose.

Press [Take Sample] to record a collected sample value in memory. Re-enter the Sample Cal.

screen and press [Start calibration] to perform a sample calibration. This updates the recorded

data.

8.5

Temperature coefficient

Simply input the solution conductivity at the reference temperature (TR) after allowing the sensor

to stabilize at elevated temperatures. The FLXA21 will calculate the temperature coefficient for

you. The ideal temperature for this calibration is the normal process value (TP).
This calibration is enabled if the Temperature Compensation is set to “TC.” (Section 7.1.4)

8.6

Temperature calibration

In order to make the most accurate measurements, it is important to have a precise temperature

measurement. This affects the display of temperature, and the output signal (when used). More

important, however, is the temperature compensation, and calibration accuracy.
The temperature of the sensor system should be measured independently with a high precision

thermometer. The value of “Measured temperature” should then be adjusted to agree with the

reading (zero offset calibration only).
For best accuracy, this should be done as near to the normal operating temperature as possible.

SC

8

This manual is related to the following products: