<1. INTRODUCTION AND GENERAL DESCRIPTION>
1-11
IM 12A01A02-01E
5th Edition : Oct. 31, 2013-00
pH/ORP Sensor module, SC Sensor module and DO Sensor module
Input parameters
Ui, Vmax
= 13.92 V
Ii, Imax
= 50 mA
Pi, Pmax
= 0.374 W
Ci
= 40 nF
Li
= 2.9 mH
Output parameters
Sensor input circuit(pH: terminals 11 through 19,
SC: terminals 11 though 16, DO: terminals 11 thourgh 18)
Uo Vt, Voc
= 11.76 V
Io, It, Isc
= 116.5 mA
Po
= 0.342 W
Co, Ca
= 100 nF
Lo, La
= 1.7mH
ISC Sensor module
Input parameters
Ui, Vmax
= 13.92 V
Ii, Imax
= 50 mA
Pi, Pmax
= 0.374 W
Ci
= 40 nF
Li
= 7.7 mH
Output parameters
Sensor input circuit (terminals 11 through 17)
Uo Vt, Voc
= 11.76 V
Io, It, Isc
= 60.6 mA
Po
= 0.178 W
Co, Ca
= 100 nF
Lo, La
= 8 mH
Model code as per certification unit
Housing assembly
Model code Enclosure type (Housing)
Display
Explosion proof type
K9698NA
Plastic (Polycarbonate) (-P) Without display (-N)
Intrinsically safe type
K9698PA
With LCD display (-D)
K9698QA
Stainless steel with
urethane coating (-U)
Without display (-N)
K9698RA
With LCD display (-D)
K9698QB
Stainless steel with epoxy
coating (-E)
Without display (-N)
K9698RB
With LCD display (-D)
K9698QC
Stainless steel (-S)
Without display (-N)
K9698RC
With LCD display (-D)
Sensor module
Model code
Sensor module type
Explosion proof type
K9698EA
pH/ORP module
Intrinsically safe type
K9698FA
SC module
K9698GA
ISC module
K9698HA
DO module
Control Drawing (ATEX and IECEx types)
Measuring module 1
Note: The measuring module on this drawing means
the sensor module on this General Specifications.
Measuring module 2
Sensor 1
Sensor 2
(*)
(Note 1)
Safety Barrier
Housing Assembly
Supply -
Supply +
→
Non Hazardous Location
Hazardous Location
←
-
+
Note: The measuring module on this drawing means the sensor module on this General Specifications.
1
PH
SC
ISC
DO
SENCOM