<13. COMMISSIONING OF DO (Dissolved Oxygen)>

13-6

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

mA

mA

mA

R(1)

R(2)

R(1)

F0503_2.ai

Return mA to Sensor1?

Yes
No

Reset redundant system

Sensor 1

Display

Sensor 2

Alive

Alive

Alive

Alive

Alive

Dead

Dead
Alive

Sensor 1 is dead.

(automatic)

Repair sensor

Redundant system

restart (manual)

If sensor 1 fails, the

output is automatically

switched to the sensor-2

value.

This display is the

example when

“Redundant” is

selected as a

process parameter.

Even if a fault on the sensor 1 is

recovered automatically after

fault detection, the output will not

be switched back automatically

and the sensor 2 value will

continue to be output.

After repairing sensor 1, reset the backup made

by the redundant system. This enables the

sensor 1 value to be incorporated in the output.

On the Reset redundant system, selection of

“Yes” makes the output return to the output of

the 1st module.

Output setup

mA

Output






Configure Hold
Reset redundant

Figure 13.2

Redundant system

Setup
Select one of the output methods: Linear and Table.

Linear:

Set the 0% and 100% values.

Table:

This allows the configuration of an output curve by 21 points (5% intervals).

(The 0% and 100% values must be entered. For the default values, see Table 13.3)

Table 13.3

Table of Default Values

%

0.000 5.000 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00

Oxygen (mg/L)

0.00

1.00

2.00

3.00

4.00

5.00

6.00

7.00

8.00

9.00 10.00

Oxygen (ppm)

0.00

1.00

2.00

3.00

4.00

5.00

6.00

7.00

8.00

9.00 10.00

Oxygen (ppb)

0.0

5.0

10.0

15.0

20.0

25.0

30.0

35.0

40.0

45.0

50.0

Oxygen (%sat)

0.0

5.0

10.0

15.0

20.0

25.0

30.0

35.0

40.0

45.0

50.0

Temperater (°C)

0.0

2.5

5.0

7.5

10.0

12.5

15.0

17.5

20.0

25.5

25.0

%

55.00 60.00 65.00 70.00 75.00 80.00 85.00 90.00 95.00 100.0

Oxygen (mg/L)

11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00

Oxygen (ppm)

11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00

Oxygen (ppb)

55.0

60.0

65.0

70.0

75.0

80.0

85.0

90.0

95.0 100.0

Oxygen (%sat)

55.0

60.0

65.0

70.0

75.0

80.0

85.0

90.0

95.0 100.0

Temperater (°C)

27.5

30.0

32.5

35.0

37.5

40.0

42.5

45.0

47.5

50.0

Burn
Select the designated output in case of a fault from among Off, Low, and High. See “13.4 Error

configuration” to set the output.

Off: Output depends on the measured value. (No fixed output is produced.)
Low: Output is fixed to 3.6 mA
High: Output is fixed to 22.0 mA.