beautypg.com

4 sample, 5 temperature coefficient, 4 sample -4 11.5 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 139: Temperature coefficient -4

background image

<11. CALIBRATION OF ISC (Inductive Conductivity)>

11-4

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

NOTE

The temperature compensation should be set to NaCl when confirming zero during air

calibration.

NOTE

The temperature compensation is not performed around zero. In this case, a warning may be

issued.

Not being wet, the sensor measures the conductivity of air, which is close to zero. Because

zero conductivity is out of the range for temperature compensation (this range is based on the

temperature and conductivity of solutions and is preset to 0.033 μS/cm at 20°C), a warning may

be issued. This is not a device error.
The conductivity of air is zero and it does not have a temperature coefficient. Execute zero

calibration regardless of the warning.

11.4 Sample

With the sensor in situ, a sample can be taken for laboratory analysis. Sample calibration records

the time and reading, and holds these in memory until the analysis has been completed. The

laboratory data can then be entered regardless of the current process value, without the need for

calculations.
When sample calibration is made with temperature compensation activated, the types of

temperature compensation for laboratory analysis equipment should be matched. Use of a

different type of temperature compensation between equipment causes an error. The standard

conductivity meter to be used should always be based on the accurate and same temperature

compensation calculation method. Yokogawa recommends that the Model SC82 pocket

conductivity meter be used for this purpose.

Press [Take Sample] to record a collected sample value in memory. Re-enter the Sample Cal.

screen and press [Start calibration] to perform a sample calibration. This updates the recorded

data.

11.5 Temperature coefficient

Simply input the solution conductivity at the reference temperature (TR) after allowing the sensor

to stabilize at elevated temperatures. The FLXA21 will calculate the temperature coefficient for

you. The ideal temperature for this calibration is the normal process value (TP). This calibration is

enabled if the Temperature Compensation is set to “TC.” (Refer to section 10.1.4)

This manual is related to the following products: