<1. INTRODUCTION AND GENERAL DESCRIPTION>

1-13

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

Control Drawing (CSA type)

Measuring module 1

Measuring module 2

(Refer to Note 7)

Sensor 1

Sensor 2

Housing Assembly

Supply -

Supply +

Non Hazardous Location

Hazardous Location

(*1)

Refer to Note

Safety Barrier

-

+

Intrinsically Safe

Group IIC, Zone 0

Class I, Division 1

Non-incendive

Class I, Division 2,

Groups A, B, C, D

2 wire analyzer

Not use safety barrier but

CSA certified equipment

use in Non-incendive

Note: The measuring module on this

drawing means the sensor module

on this General Specifications.

Electrical parameters (Intrinsically safe)

Housing Assembly

Supply and output circuit (terminals + and -)

Ui(Vmax)=30V, Ii(Imax)=100mA, Pi(Pmax)=0.75W, Ci=13nF, Li=0mH

Measuring module input circuit (CN2 or CN3 on Back board)

Uo(Vt,Voc)=13.65V, Io(It,Isc)=50mA, Po=0.372W, Co(Ca)=80nF, Lo(La)=7.7mH

pH module, SC module and DO module

Ui(Vmax)=13.92V, Ii(Imax)=50mA, Pi(Pmax)=0.374W, Ci=40nF, Li=2.9mH

Sensor input circuit (terminals 11 through 19)

Uo(Vt,Voc)=11.76V, Io(It,Isc)=116.5mA, Po=0.3424W, Co(Ca)=100nF, Lo(La)=1.7mH

ISC module

Ui(Vmax)=13.92V, Ii(Imax)=50mA, Pi(Pmax)=0.374W, Ci=40nF, Li=7.7mH

Sensor input circuit (terminals 11 through 17)

Uo(Vt,Voc)=11.76V, Io(It,Isc)=60.6mA, Po=0.178W, Co(Ca)=100nF, Lo(La)=8mH

Installation requirements between housing assembly and safety barrier

Uo≤Ui Io≤Ii Po≤Pi Co≥Ci+Ccable Lo≥Li+Lcable

Voc≤Vmax Isc≤Imax Ca≥Ci+Ccable La≥Li+Lcable

Uo, Io, Po, Co, Lo, Voc, Isc, Ca and La are parameters of barrier.

Electrical Parameters (Nonincendive)

Housing Assembly

Supply and output circuit (terminals + and -)

Ui(Vmax)=30V, Ci=13nF, Li=0mH

Measuring module input circuit (CN2 or CN3 on Back board)

Uo(Vt,Voc)=13.65V, Io(It,Isc)=50mA, Co(Ca)=80nF, Lo(La)=7.7mH

pH module, SC module and DO module

Ui(Vmax)=13.92V, Ci=40nF, Li=2.9mH

Sensor input circuit (terminals 11 through 19)

Uo(Vt,Voc)=11.76V, Io(It,Isc)=116.5mA, Co(Ca)=4uF, Lo(La)=4.5mH

ISC module

Ui(Vmax)=13.92V, Ci=40nF, Li=7.7mH

Sensor input circuit (terminals 11 through 17)

Uo(Vt,Voc)=11.76V, Io(It,Isc)=60.6mA, Co(Ca)=4uF, Lo(La)=19mH

Note for Intrinsically Safe Installation:

1: In any safety barrier used, the output current must be limited by a resistor “R” such that Io=Uo/R or Isc=Voc/R.

2: The safety barrier must be CSA certified.

3: Input voltage of the safety barrier must be less than 250Vrms/Vdc

4: When using non isolation barrier connect (*1) to IS earthing system.

5: pH and SC Sensor(s) are of a passive type to be regarded as ‘simple apparatus’ same as 06ATEX0218X, 06ATEX0219,

IECEx KEM 06.0052X, FM3028779, 06ATEX0220X, 06ATEX0221, IECEx KEM 06.0053X or the one individually certified with

relevant parameters.

6: ISC Sensor(s) are ISC40S of 00ATEX1067X or the one individually certified with relevant parameters.

7: DO Sensor(s) are of a passive type to be regarded as ‘simple apparatus’ or the one individually certified with relevant

parameters.

8: Measuring module 2 may not mounted. As for ISC module, only one can be mounted.

9: Installation should be in accordance with Canadian Electrical Code Part I and Local Electrical Code.

10: Do not alter drawing without authorization from CSA.

Note for Nonincendive Installation:

1: The parameter for sensor input circuit must be taken into account when installed.

2: Installation should be in accordance with Canadian Electrical Code Part I and Local Electrical Code.

3: Do not alter drawing without authorization from CSA.