Page 107: 7 calculated data setup, Auto return -15, Adjust contrast -15, Monitor display -15, Calculated data setup -15, Sc 7

background image

<7. COMMISSIONING OF SC (Conductivity)>

7-15

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

7.6.3

Auto Return

When no operation is performed for the time set in “Auto Return”, the display returns to the

Monitor display (or to the Main display when the MONITOR display is disabled) and the analyzer

returns to a normal measuring mode. (When the Trend display is selected, the Auto Return

doesn’t work.)
Select the time from among Disable, 10 min, and 60 min. When the Auto Return function is not

used, select “Disable.”

NOTE

A default is “10 min”. When maintenance like a calibration that may take much time is performed,

“60 min” or “Disable” is recommended to be selected.

7.6.4

Adjust contrast

The LCD screen contrast can be adjusted.
Pressing the ▲▼ keys adjusts the contrast in 11 levels from +5 to –5 (including the default value

of “0”).

7.6.5

MONITOR display

Select “Enable” so that the Monitor display becomes available. A default is “Enable”.
During Hold condition and a warning/fault condition, the Main display or the Home display is

displayed to indicate the condition.

7.7

Calculated data setup

On this setup, parameters can be set for calculated data.
When two modules are installed on the instrument, the calculated data can be set for conductivity

values or resistivity values measured by two sensors.
On the Calculated data setup screen, select one kind of calculations as shown on the Table 7.4.
For resistivity, select “Differential” or “Average”.

Table 7.4

Calculation

Function

Arithmetic expression

Indication *

Differential

(value of the sensor 1) – (value of the sensor 2)

Diff

Average

(value of the sensor 1 + value of the sensor 2) / 2

Ave

Ratio

(value of the sensor 1) / (value of the sensor 2)

Ratio

Passage(%)

(value of the sensor 2) / (value of the sensor 1) x 100

Pass

Rejection(%)

(value of the sensor 1 – value of the sensor 2) / (value of the sensor 1) x 100

Reje

Deviation(%)

(value of the sensor 2 – value of the sensor 1) / (value of the sensor 1) x 100

Dev

pH calc.(VGB)

pH = 8.6 + log {(value of the sensor 1) – (value of the sensor 2) /3}

VGB

*

Parameter symbol at the bar display of analog output which is displayed at the bottom of the Main display or the Home display

when a calculated data is selected as a process parameter on the “mA (Output)” setup screen.

To display the calculated result, select “Calculated” on the Dual display screen. (Section 7.6.1.)

(The setting on the “Individual display” has no effect.)
While displaying a calculated data, measurement value of the sensor 1 or sensor 2 can be

checked on the each sensor’s display with pressing

Sensor

1

or

Sensor

2

at the lower right on the Main

display.

SC

7