Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 95
Section 12: Specifications
05/11 96
51209 Revision 2
12.2 General Data
The following characteristics are valid after a warm-up time of at least 60 minutes (reference
temperature: 23°C).
12.2.1 Power Supply
Supply voltage
24 VDC
+/- 20 %
Power
consumption
6 Watts
Power
consumption
with fieldbus option
8 Watts
12.3 Effect of Ambient Conditions
12.3.1 Environmental Conditions
Temperature range
Reference temperature
23 °C
Ambient temperature for
operation
10,000 e
-10... +40 °C
‘not legal for
trade’
-10... +50 °C
Power-on temperature
0... +50 °C
Storage/transport
-40... +70 °C
Humidity
< 95 %, no condensation, (acc. to IEC 68-2)
Protection type to DIN 40050
IP30
12.3.2 Electromagnetic Compatibility (EMC)
All data in compliance with EN 61326 industrial area
Housing
Radio frequency electromagnetic fields (80 – 1000 MHz) EN 61000-4-3
10 V/m
Radio frequency electromagnetic fields (1.4 – 2.0 GHz)
EN 61000-4-3
3 V/m
Radio frequency electromagnetic fields (2.0 – 2.7 GHz)
EN 61000-4-3
1 V/m
Electrostatic discharge (ESD)
EN 61000-4-2
6 / 8 kV
Signal and
Fast transients (burst)
EN 61000-4-4
1 kV
control lines Peak voltages (surge) 1.2/50 µs
EN 61000-4-5
1 kV
Conducted disturbances by radio frequency (0.15 –
80 MHz)
EN 61000-4-6
10 V
Mains input
Fast transients (burst)
EN 61000-4-4
2 kV
Peak voltages (surge) 1.2/50 µs
EN 61000-4-5
1 / 2 kV
Conducted disturbances by radio frequency (0.15 –
80 MHz)
EN 61000-4-6
10 V
12.3.3 RF Interference Suppression
Electromagnetic emission
In acc. with EN 61326, limit value class A, industrial area