8 global spm variables – Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 90
Section 8: Global SPM Variables
05/11 91
51209 Revision 2
8 Global SPM Variables
For communication via OPC, the following variables are available:
BOOL R
Digital input 1
X 0
BOOL R
Digital input 2
X 1
BOOL R
Digital input 3
X 2
BOOL W
Digital output 1, transparent mode
X 8
BOOL W
Digital output 2, transparent mode
X 9
BOOL W
Digital output 3, transparent mode
X 10
BOOL R
Output Limit 1
X 16
BOOL R
Output Limit 2
X 17
BOOL R
Output Limit 3
X 18
BOOL R
ADC error
X 32
BOOL R
Above Max
X 33
BOOL R
Overload, above (Max + plus the range that is still permissible)
X 34
BOOL R
Below zero
X 35
BOOL R
Zero +/- 1/4 d
X 36
BOOL R
Within the zero setting range
X 37
BOOL R
The weight is stable
X 38
BOOL R
Below zero or above Max
X 39
BOOL R
Load cell measuring signal negative
X 40
BOOL R
Load cell measuring signal > 36 mV
X 41
BOOL R
Internal arithmetic error; CAL data are perhaps faulty
X 42
BOOL R
ADC error, overload
X 43
BOOL R
Command error
X 48
BOOL R
Command active
X 49
BOOL R
Power fail signal
X 50
BOOL R
ADC test active
X 56
BOOL R
Calibration active
X 57
BOOL R
Instrument is tared
X 58
BOOL R/W Read / write marker bit 1
X 64
BOOL R/W Read / write marker bit 2
X 65
BOOL R/W Read / write marker bit 3
X 66
BOOL W
Switch to net weight
X 72
BOOL W
Set the instrument to zero
X 112
BOOL W
Set tare of the instrument
X 113
BOOL W
Reset the tare of the instrument
X 114
BOOL W
Start the ADC test
X 115
BOOL W
Finish the ADC test
X 116
BOOL W
Reset the power fail signal
X 117
BOOL W
Set the fixtare value as tare
X 118
BOOL W
Store the current gross weight in the fixtare memory (D31)
X 119
BOOL W
Start printing
X 120
BOOL W
Reset error
X 121