Section 5: Commissioning

05/11 45

51209 Revision 2

5.4.19 Parameter Input


Open the menu via

-[Calibration]-[Param].

WP A/Calibration

Measuretime

160 ms

Digital filter

bessel

External supply

8 - 12 VDC

Fcut

3.00 Hz This line is shown only if the filter is on.

Test mode

absolute

W & M

none

Standstill time

0.50 s

Standstill range

1.00 d


[Measuretime]
Select the measuring time; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms.
In 'legal-for-trade' mode select <= 1 s.

[Digital filter]
The digital filter can be switched on only with the measuring time set to <= 160 ms.
Select the filter characteristic [off], [bessel], [aperiod.], [butterw.], [tscheby.].

[External supply]
With external load cell supply (e.g. 7.5 VDC via 1626/60 with MX8 = closed), [below 8 VDC] can be
selected to adapt the sense voltage monitoring to the lower supply voltage.

[Fcut]
Enter the cut-off frequency for the filter (0.1 – 80 Hz); the setting is dependent on the measuring
time.

[Test mode]
With [Absolute], the test value is calculated when the test is called. With [relative], the deviation
from the initially stored test value is displayed; see Chapter 5.4.17.

[W & M]
Setting for 'legal-for-trade' mode; select [none], [OIML], [NTEP] (for USA) or [NSC] (for Australia),
see Chapter 5.4.20.

[Standstill time]
Settings [Standstill range] and [Standstill time] are required for determining the mechanical stability
of the scale. Input in seconds; permissible range 0.01 bis 2 sec. (max. measuring time x 32). If 0 is
set, stability is not checked. The stability time must not be less than the measuring time and not
greater than 32 times the measuring time.

[Standstill range]
The scale is stable as long as any changes in the weight value are within this range; permissible
range: 0.01d to 10.00d. In 'legal-for-trade' mode select <= 1 d.