Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 44
Section 5: Commissioning
05/11 45
51209 Revision 2
5.4.19 Parameter Input
Open the menu via
-[Calibration]-[Param].
WP A/Calibration
Measuretime
160 ms
Digital filter
bessel
External supply
8 - 12 VDC
Fcut
3.00 Hz This line is shown only if the filter is on.
Test mode
absolute
W & M
none
Standstill time
0.50 s
Standstill range
1.00 d
[Measuretime]
Select the measuring time; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms.
In 'legal-for-trade' mode select <= 1 s.
[Digital filter]
The digital filter can be switched on only with the measuring time set to <= 160 ms.
Select the filter characteristic [off], [bessel], [aperiod.], [butterw.], [tscheby.].
[External supply]
With external load cell supply (e.g. 7.5 VDC via 1626/60 with MX8 = closed), [below 8 VDC] can be
selected to adapt the sense voltage monitoring to the lower supply voltage.
[Fcut]
Enter the cut-off frequency for the filter (0.1 – 80 Hz); the setting is dependent on the measuring
time.
[Test mode]
With [Absolute], the test value is calculated when the test is called. With [relative], the deviation
from the initially stored test value is displayed; see Chapter 5.4.17.
[W & M]
Setting for 'legal-for-trade' mode; select [none], [OIML], [NTEP] (for USA) or [NSC] (for Australia),
see Chapter 5.4.20.
[Standstill time]
Settings [Standstill range] and [Standstill time] are required for determining the mechanical stability
of the scale. Input in seconds; permissible range 0.01 bis 2 sec. (max. measuring time x 32). If 0 is
set, stability is not checked. The stability time must not be less than the measuring time and not
greater than 32 times the measuring time.
[Standstill range]
The scale is stable as long as any changes in the weight value are within this range; permissible
range: 0.01d to 10.00d. In 'legal-for-trade' mode select <= 1 d.