Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 47
Section 5: Commissioning
05/11 48
51209 Revision 2
5.5 Error Messages
5.5.1
Measuring Circuit Error Messages
The weighing electronics can generate error messages, which are output on the weight display.
VNC Text Cause
Arith err
Internal arithmetic overflow (faulty calibration values)
Overload Input voltage is higher than Max + (x d)
No EOC Input signal is higher than the permissible range of 36 mV. However,
the trouble cause can also be due to an error in the analog section, to
a defective load cell or to a cable break
No sense
voltage
Sense line or supply line is interrupted or connected with wrong
polarity or sense voltage is low
Negative
Input
Input voltage is negative (e.g. wrong polarity of load cell signal or
supply voltage)
5.5.2
Testing the Measuring Circuit
A simple test with the load cells connected can be carried out using a multimeter (not with external
supply or intrinsically safe load cell interface):
12 V +/- 0.8 V
(symmetrical to housing GND)
12 V +/- 0.8 V
(symmetrical to housing GND)
0 - 12 mV @ LC with 1.0 mV/V
0 - 24 mV @ LC with 2.0 mV/V
Supply voltage
Sense voltage
Measuring voltage