Section 5: Commissioning

05/11 47

51209 Revision 2

5.4.20 Legal-for-Trade Operation


Under

-[Calibration]-[Param]-[W

&

M] you can choose between: [none] and a legal-for-trade mode

[OIML], [NTEP] or [NSC].

[none]

[OIML]

[NTEP]

[NSC]

Gross weight display

Min. meas. signal

0.125 mV/V

at 30000 d

0.125 mV/V

at 3000 e

0.125 mV/V

at 3000 e

0.125 mV/V

at 3000 e

0.25 mV/V
at 60000 d

0.25 mV/V

at 6000 e

0.25 mV/V

at 6000 e

0.25 mV/V

at 6000 e

0.42 mV/V
at 10000 e

0.42 mV/V
at 10000 e

0.42 mV/V
at 10000 e

If legal-for-trade operation is selected, the parameters (zero tracking etc.) must be set accordingly;
they are not checked. The relevant CAL switch (see Chapter 5.1.1) must be sealed in the closed
position.


5.4.21 Multiple Range Scale/Multi-Interval Scale

Range selection is controlled by three parameters under

-[Calibration]-[Param].

WP A/Calibration

Range mode

Multiple range Select [Multiple range] or [Multi-interval]

Range limit 1

1000 kg Switch point from range 1 to 2

Range limit 2

2000 kg Switch point from range 2 to 3


Multiple range scale (cl. III, or cl. I and II single range scale with variable scale interval)
With [Range mode] = [Multiple range], the scale has up to 3 ranges with different resolutions.
The switch points [Range limit 1] and [Range limit 2] are the range limits. As soon as the gross
weight exceeds range 1, the next higher range with the next higher interval becomes valid (1->2->5-
>10->20->50->100->200). When reducing the weight, the interval of the previous range is kept.
When the gross weight is <= 0.25 d of range 1 and the scale is stable and not tared, the scale
returns to range 1 with the corresponding interval.

Multi-interval scale (cl. III, or cl. I and II single range scale with variable scale interval)
With [Range mode] = [Multi-interval], the scale has up to three ranges with different resolutions.
Each range has the corresponding interval. Unlike [Multiple range], switching the interval is also
triggered by weight reduction; i.e., when the weight drops below the range limits.

During calibration, the multiple range/multi-interval function is always switched off.
The weight display header (VNC) includes the current range (R1, R2, R3), Max, Min and d (or e
with legal-for-trade instruments) (Example: multiple range scale in range 2):


The ranges are marked by points on the left side of the display.

Range 1

Range 2

Range 3