Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 19
Section 3: Installing the Instrument
05/11 20
51209 Revision 2
3.2.4
Connecting PR 6221 Load Cells
See PR 6021/08, -/68 operating manual.
3.2.5
External Load Cell Supply
The internal load cell supply voltage of PR5220 (V+, V-) is not connected.
The common line of the symmetrical external supply must be connected to the same terminal of PR
5220 as the shield of the load cell / extension cable.
Specification of external supply: +/- 6 VDC + 5%, -30%; max ripple. 50 mVpp; max. asymmetry +/-
3%. An external supply voltage smaller than 8 VDC (+/- 4 VDC) must be set under
-[Calibration]-
[Param]-[External supply].
3.2.6
Connection via PR 1626/60
Connect the instrument to PR 1626/60 as described below. For additional connections, refer to the
PR 1626/60 manual. The internal load cell supply voltage of PR5220 (V+, V-) is not connected.
If MX8 is closed, [below 8VDC] must
be set under
-[Calibration]-[Param]-
[External supply].