Section 3: Installing the Instrument

05/11 20

51209 Revision 2

3.2.4

Connecting PR 6221 Load Cells

See PR 6021/08, -/68 operating manual.

3.2.5

External Load Cell Supply


The internal load cell supply voltage of PR5220 (V+, V-) is not connected.
The common line of the symmetrical external supply must be connected to the same terminal of PR
5220 as the shield of the load cell / extension cable.

Specification of external supply: +/- 6 VDC + 5%, -30%; max ripple. 50 mVpp; max. asymmetry +/-
3%. An external supply voltage smaller than 8 VDC (+/- 4 VDC) must be set under

-[Calibration]-

[Param]-[External supply].

3.2.6

Connection via PR 1626/60


Connect the instrument to PR 1626/60 as described below. For additional connections, refer to the
PR 1626/60 manual. The internal load cell supply voltage of PR5220 (V+, V-) is not connected.

If MX8 is closed, [below 8VDC] must
be set under

-[Calibration]-[Param]-

[External supply].