Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 50
Section 5: Commissioning
05/11 51
51209 Revision 2
5.6.3
Operating Parameters
Define the basic operating parameters under
-[
Operating parameter
].
Setup/Operating parameter
AccessCode
0
SetTareKey
tare & reset tare
SetZeroKey
only when not tared
[AccessCode]
The access code can be used to protect the [Setup] from unauthorized operation. Enter a number
with up to 6 digits. As long as you are in this menu, the value can be overwritten as required. When
[AccessCode] is set to 0, no access code ist prompted.
[SetTareKey]
The SetTare function can be configured (VNC operation / Internet browser).
[tare & reset tare]: the scale is tared if it was not tared previously; otherwise, tare is reset.
[tare & tare again]: when you press the [Tare] key, the current value is stored in the tare memory
and a net weight of 0 is displayed.
[disabled]: The tare key has no function.
[SetZeroKey]
The SetZero function (VNC operation / Internet browser) can be limited to gross mode with [only
when not tared] or switch the scale to gross mode automatically with [reset tare on zeroset]. If the
zero-setting key with these settings has no effect, the configured zero-setting range (around the
zero-point set with the dead load) is already utilized due to a previous zero-setting operation and/or
automatic zero setting.
[disabled]: The key has no function.
Save changes ?
To close the menu, press
. Press [Yes] to save the data,
or [No] to close the menu without changing data.