Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 32
Section 5: Commissioning
05/11 33
51209 Revision 2
5.3.4
Setup Function (VNC)
Press
to configure the main operating parameters. The instrument configuration depends on the
application and the plug-in cards installed.
Calibration is in a simple dialogue. Compliance with the relevant (verification) standards must be
checked by the person commissioning the instrument or the verification officer. To protect the
calibration data from overwriting, close the CAL switch on the back panel of the instrument. On
legal-for trade instruments, the CAL switch must be sealed in the closed (write-protected) position;
see Chapter 5.1.1.
5.3.5
Setup Menu
- Serial ports parameter
- Remote display
- SMA
not used
232
- Param
Builtin RS232, Assigned to, Protocol, Baud rate., Bits, Parity,
Stop bits,
- Operating parameter
- AccessCode
Access code for changes
- SetTareKey
Tare&reset tare, tare&tare again, disabled
- SetZeroKey
Only when not tared, reset tare on zeroset , disabled.
- Fieldbus parameter
Only for PR 5220/01 and /04, see Chapter 5.6.4
- Network parameter
Hardware address (read only), Instrument name, IP address,
Subnet mask, Standard gateway, VNC-Client (access restriction)
- Calibration
New, Modify, Param, see Chapter 5.3.6
-Limit parameter
- Limit 1/2/3 on/off
Limit 1/2/3 'on'/ 'off', Action, Condition; see Chapter 5.7.5
-Digital I/O parameter
- Output 1/2/3
Configuring outputs; see Chapter
- Limit 1/2/3 on/off
Inputs, Action, Condition; see Chapter 5.7.4
- Analog output parameter
Gross, Net, Select, Transparent; see Chapter 5.8