Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 20

Section 3: Installing the Instrument
05/11 21
51209 Revision 2
3.2.7
RS-422/485 Interface
 
The interface is intended for connecting a PR 1627 type remote display, a PR 1628 terminal or a 
PC for data transmission using the SMA protocol. 
 
Connection method:
4-pin plug-in terminal block
Number of channels / 
type: 
1 RS-422 / RS-485 , full / half duplex
Transfer rate (Bits/s):
300, 600, 1200, 2400, 4800, <9600>, 
19200 
Bits / stop bits:
<8 / 1> or 7 / 1
Parity:
Signals:
RxA (R-), RxB (R+), TxA, TxB
Electrical isolation:
Yes
Cable length:
Max. 1000m
Cable type:
Shielded twisted pair (e.g. LifYCY 
2x2x0.20) 
<...> = default settings (factory settings) 
 
 
Connection of a PR 1627 remote display or of a PR 1628/00, -/60 or -/24 terminal
(over the PR 1604 interface)
 
 
 
The following operations are possible from the 
connected terminal / display: 
 
Test 
Set tare 
Reset tare 
Set zero 
 
 
 S300 
ON <-----
OFF ----->
2-wire connection
(d) 4-wire connection
(d) Receiver terminating resistor 'ON'
Receiver terminating resistor 'OFF'
(d) Transmitter (rec.) terminating resistor 
'ON' 
Transmitter (rec.) terminating resistor 'OFF'
 Calibration data and parameters 
secured 
(d) Calibration data and parameters not 
secured 
 
(d) - default = factory setting 
 
 
