Section 5: Commissioning
05/11 49
51209 Revision 2
5.6 Configuring General Parameters
The configuration of parameters which are not related to the weighing electronics is divided into
several sections (see Chapter 5.3.5).
5.6.1
Serial Ports
To configure the serial interfaces, select
-[Serial ports parameter].
Setup
Serial ports parameter
Select [Serial ports parameter] with
Operating parameter
and
Setup/Serial ports
Remote display
Builtin RS485
SMA
none
Param
[Remote display]
Select the serial interface to which the remote display is connected and then select [Param] to
define the [Baudrate].
Setup/Serial ports/Builtin RS485
Assigned to
Remote display
Protocol
Remote display
Baudrate
9600 bd Select [Baud rate] with
Bits
7
and set the baud rate with
Parity
even
Stopbits
1
Mode
single transmitter
You can only set the baud rate to 300, 600, 1200, 2400, 4800 or 9600. The other parameters
(displayed in light-gray) cannot be changed.
If only 1 instrument is connected to a PR1627 or PR1628 (normal case), [Mode] must be set to
[single transmitter].
Save settings