Section 5: Commissioning

05/11 39

51209 Revision 2

5.4.9

Determining the Dead Load

WP A/Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

0.00 kg

0.000000

Max at

3000.00 kg

1.000000

Not calibrated

Sensitivity

833.33

4.000000

by load

by mV/V

Test


To use the empty scale/hopper as dead load (normal case):
- discharge the scale/hopper
- press [by load]

Set deadload

After

or

confirmation of the change is displayed:


If the mV/V value of the dead load was calculated, or if it is known from the previous calibration, the
value can be overwritten by pressing [by mV/V].

WP A/Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

0.00 kg

0.000000

Max at

3000.00 kg

1.000000

Calibrated at

3000.00 kg

1.000000

Sensitivity

833.33

4.000000

by load

by mV/V

Test

5.4.10 Possible error messages

Set deadload failed

above physmax

The dead load entered in mV/V plus maximum capacity in mV/V is
higher than 3 mV/V ( = 36 mV).

Set deadload failed

no standstill

The scale is not stable.
Remedial action: Check the mechanical function of the scale; adapt the
filter setting; reduce the resolution; if necessary, adapt the stability
conditions.

Set deadload failed

deadload < -0.1mV/V

Measurement signal is negative (load cells connected with wrong
polarity or defective) when determining the dead load with [by load].

Set deadload failed

overflow in arithmetics

The dead load entered in mV/V is higher than 5mV/V.