Fairbanks X SERIES PR 5220 ETHERNET TRANSMITTER User Manual
Page 35

Section 5: Commissioning
05/11 36
51209 Revision 2
5.4.4
Selecting the Calibration Mode
You can choose between [New] and [Modify] with the softkeys: 
 
New
Modify
Param
 
 
5.4.5
New Calibration
Open the menu via
-[Calibration].
When you press [New], the data is set to default first and calibration is started. 
 
SPAN and deadload
will be reset
You are prompted to confirm. 
Press [Continue] for the default settings, or [Cancel] to cancel the 
selection. 
 
Default settings with [New]: 
WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.000000
Max at
1.000000
Not calibrated
Sensitivity
833.33
4.000000
Test
 
 
5.4.6
Changing the Calibration
[Modify] may be used only for minor changes (e.g. changing the dead load, adapting mV/V 
values for dead load and/or Max); otherwise, always use [New]. 
 
Open the menu via 
-[Calibration]-[Modify].
WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
1.07 kg
0.000358
Max at
3000.00 kg
1.000000
Calibrated at
3000.00 kg
1.000000
Sensitivity
833.33
4.000000
by load
by mV/V
Test
 
For setting a new value for Dead load, press 
to select [Deadload] and either enter a new
value with [by mV/V] or discharge the scale/hopper and press [by load].
Exit calibration
without CalcTest?
When closing the menu with
you are prompted whether the menu
should be closed without calculation of the test value: 
Reply [Yes] to close the menu. 