Section 3: Installing the Instrument

05/11 19

51209 Revision 2

3.2 Connections

3.2.1

Load Cell Connection

The cable colors shown in this chapter are applicable to the PR 62XX series load
cells.
Before connecting other types, please, carefully follow the information related to the
assignment of load cell / platform cable colors.


3.2.2

Connection Using 6-Wire Technology

See also label on the housing outside (Chapter 2.3) and manual of the cable junction box.

Terminal

Description

SIGN. M+

SIGN. M-

SENSE S+

SENSE S-

LC SUPPLY

V+

LC SUPPLY

V-

+ signal / LC output
- signal / LC output
+ sense
- sense
+ supply / excitation
- supply / excitation

Recommendations:

- Install cable in steel pipe connected to earth potential

- Min. distance to high-voltage cables: 1m


Load cell supply circuit:

The load cell supply voltage is fixed to 12VDC and protected against short circuit
Load resistance of load cells > 75 ohms, e.g. 8 load cells of 650 ohms each


3.2.3

Connection of a Load Cell in 4-Wire Technology

Note that links between SENSE S+ and LC SUPPLY V+ and between SENSE S- and LC SUPPLY
V- directly at the transmitter must be provided.

Terminal

Description

SIGN. M+

SIGN. M-

SENSE S+

SENSE S-

LC SUPPLY

V+

LC SUPPLY

V-

+ signal / LC output
- signal / LC output
+ sense
- sense
+ supply / excitation
- supply / excitation