39

Signet 8900 Multi-Parameter

Adjust LCD contrast: 4 settings.

If two or more measurements of the same type are present, select a functional relationship

between them: Sum Difference Ratio % Reject % Passage % Recovery (A, B, or C)

Power 2 Power 4 None

Enter notes such as:

next commissioning date, scheduled service, Tech Support telephone number, etc.

The standard password is "UP-UP-UP-DOWN" Select Enhanced password to set a

custom 4-digit code.

If "Enhanced" password is selected, set any 4-digit numerical code: 0000 (default) to

9999. If the enhanced password is lost, see section 14.1 below.

Select the type of data transfer you will use to clone the setup from one 8900 into another.
Refer to Appendix E for detailed instructions. Digital (S

3

L) or COMM.

Select whether this 8900 will send setup information to another unit, or will receive the

setup from another unit. SEND or RECEIVE.

Select "YES" to begin clone process.

Enable Auto Scroll for measurement Display screens in View.

Set the scroll time interval for each measurement display screen. Minimum 1 second,

maximum 99 seconds. Press any key to interrupt autoscroll for 20 seconds, or the scroll

time, whichever is greater.

14. Options

14.1

Password Security

The standard password is the UP-UP-UP-DOWN keys, pressed in sequence. This access code is designed to protect the 8900 from
unintentional changes. It is best suited for systems where a group of people need to be able to change settings.

Enhanced Password
The enhanced password default setting is 0-0-0-0, adjustable to any 4-digit numerical code up to 9999. Using a personal password
provides the maximum degree of security.

Enhanced Password Reset Procedure:
The following procedure will reset the Enhance Password to the factory default code of (0-0-0-0).

1. Disconnect

power.

2. Remove the I/O Module (remove terminals from this and all other plug-in modules installed, then the rear cover).
3. Cycle power to the unit (apply power for at least 15 seconds, then disconnect power again)
4. Install the I/O Module, the rear panel, and terminals as applicable.
5. Reconnect power to the unit.

Completely clears any characters in the MEMO fi eld.

Change the language for all menus in the 8900. Options are:

English, Italian, German, French, Spanish, Portuguese

Select "." (Decimal point) or "," (Comma) to separate fractional portion of values.

Example: 123.45 or 123,45

Select the two measurements to be used for this function.

Password Type:

Standard

Select Enhanced

Password >

Contrast:

3

>

Auto Scroll:

Off >

Scroll Time:

4 secs >

Func. 1 Type:

Ratio

>

Func. 1 Source:
ORP1-ORP2 >

Clone Port:

S3L >

Clone Operation:

Send >

Clone

Instrument: >

Memo:

gfsignet.com >

Clear Memo?

>

Choose language:
English >

Separator:

.

>