beautypg.com

GF Signet 8450 Pressure Transmitter User Manual

GF Signet Tools

background image

CAUTION!

• Remove power to unit before wiring

input and output connections.

• Follow instructions carefully to avoid

personal injury.

Signet 8450-3 Pressure Transmitter

Contents

1. Installation
2. Specifi cations
3. Electrical Connections
4. Menu Functions

Pres1: 4.5

psi

Pres2: 2.3

psi

ENTER

Signet Pressure

Transmitter

Output -

Output +

System Pwr
Loop -

System Pwr
Loop +

2

1

4

3

Sensr Gnd

(SHIELD)

Sensr IN

(RED)

Sensr V+

(BLACK)

7

6

5

8052

92 mm
(3.6 in.)

97 mm

(3.8 in.)

56 mm
(2.2 in.)

41 mm
(1.6 in.

)

Optional
Rear
Cover

SIDE VIEW
Field Mount

96 mm

(3.8 in.)

96 mm
(3.8 in.)

96 mm

(3.8 in.)

quick-clips

gasket

panel

terminals

mounting

bracket

latch

Panel Mount

SIDE VIEW

FRONT VIEW

Field Mount &

Panel Mount

Panel Mount
Installation Detail

107 mm (4.2 in.)

42 mm

(1.7 in.)

65 mm

(2.5 in.)

1. Installation

ProcessPro transmitters are available in two styles: panel mount and fi eld mount. The panel mount is supplied with the necessary
hardware to install the transmitter. This manual includes complete panel mounting instructions.
Field mounting requires one of two separate mounting kits. The 3-8052 integral kit joins the sensor and instrument together into a
single package. The 3-8050 Universal kit enables the transmitter to be installed virtually anywhere.
Detailed instructions for integral mounting or other fi eld installation options are included with the 3-8052 Integral kit or the 3-8050
Universal kit.

1.1 Panel

Installation

1. The panel mount transmitter is designed for installation using a ¼ DIN Punch. For manual panel cutout, an adhesive template is

provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.

2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE.
Press quick-clips outward and remove.

3-8450.090-3 Rev. J 01/13 English

*3-8450.090-3*

English

Open-collector output:

Hi, Lo, Off, Programmable:

• Optically isolated, 50 mA max. sink, 30 VDC max. pull-up

voltage

• Hysteresis:

User

adjustable

Environmental
• Operating temperature: -10 °C to 70 °C (14 °F to 158 °F)
• Storage temperature:

-15 °C to 80 °C (5 °F to 176 °F)

• Relative humidity:

0 to 95%, non-condensing

• Maximum altitude:

2000 m (6562 ft)

• Insulation

category:

II

• Rating:

NEMA 4X/IP65 front

Standards and Approvals
• CE, UL listed
• Immunity: EN50082-2
• Emissions:

EN55011 Class B

• Manufactured under ISO 9001 for Quality, ISO 14001

for Environmental Management and OHSAS 18001 for
Occupational Health and Safety.

Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions:

(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.

China RoHS (Go to www.gfsignet.com for details)

2. Specifi cations

General
Compatibility: Signet 2450 Pressure Sensor
Accuracy: ±1% of full scale
Enclosure:
• Case:

PBT

• Panel case gasket: Neoprene
• Window:

Polyurethane coated polycarbonate

• Keypad:

Sealed 4-key silicone rubber

• Weight:

Approx. 325 g (12 oz.)

Display:
• Alphanumeric

2 x 16 LCD

• Update rate:

1 second

• Contrast:

User selected, 5 levels

Electrical
Power:

12 to 24 VDC ±10%, regulated,

60 mA max current

Sensor Input:
• Range:

0-250 psig, 0-17 bar, 0-1700 kPa

Current output:
• 4 to 20 mA, isolated, fully adjustable and reversible
• Max loop impedance:

50 Ω max. @ 12 V,

325

Ω max. @ 18 V,

600

Ω max. @ 24 V

• Update rate:

100 ms

• Accuracy:

±0.03

mA