Signet 8900 Multi-Parameter

Select measurement type or derived function to be assigned to this output.

If I/O Module with Voltage Output is installed: Select 0-to-5 VDC or 0-to-10 VDC.

If I/O Module with Voltage Output is installed: Enter the low-end setpoint = 0 VDC.

If I/O Module with Voltage Output is installed: Enter the high-end setpoint = 5 or 10 VDC.

Adjust the output voltage at low-end setpoint: 0 to 0.5 VDC.

Adjust the output voltage at high-end setpoint: 4.5 to 5.5 VDC or 9.5 to 10.5 VDC.

If Current output card is installed: Set the low-end setpoint = 4 mA.

If Current output card is installed: Set the high-end setpoint = 20 mA.

Adjust the output current at low-end setpoint: 3.8 to 5.0 mA.

Adjust the output current at high-end setpoint: 19.0 to 21.0 mA.

Output 1 Source:

Flow 1 >

Output 1 Range:

0-5Vdc >

Output 1 0Vdc:

+25.000 GPM >

Output 1 5Vdc:

+25.000 GPM >

Adjust 1 Low:

0 V >

Adjust 1 High:

5 V >

Output 1 4mA:

+25.000 GPM >

Output 1 20mA

+25.000 GPM >

Adjust 1 Low:

4.00 mA >

Adjust 1 High:

20.00 mA >

Test Output 1:

>

Manually adjust analog output to test operation.

If 4 to 20 mA, test limits are 3.8 mA to 21.0 mA.

If 0 to 5 VDC, test limits are 0.00 to 5.50 VDC.

If 0 to 10 VDC, test limits are 0.00 to 10.50 VDC.

12. Output