GF Signet Wall Mount Accessory Kit User Manual
Signet wall mount accessory kit instruction sheet, Installation instructions, Description 2. specifications
Signet Wall Mount Accessory Kit Instruction Sheet
3-9900.095 Rev. A 03/12 English
*3-9900.095*
The Wall Mount Accessory (159 001 700) is a compact enclosure that works with the Signet 9900 Transmitter as well as other ¼ DIN
size instruments including the Signet 8900 Controller, ProcessPro and ProPoint instruments. It is equipped with a ¼ DIN cutout at the
front for the instrument (sold separately) and two ½” holes at the base for the conduit and sensor wiring. Mounting hardware and liquid
tight connectors are included for a quick and easy installation.
3. Installation Instructions:
1. Mount the Wall Mount Accessory to your wall location using the Accessory’s mounting
hardware or brass inserts (not shown).
2. Install liquid tight connectors through the Accessory’s conduit holes to suit your
application.
3. Place gasket on instrument.
4. Slide instrument from front through cutout on door into Wall Mount Accessory opening.
5. Slide mounting bracket over back of instrument until bracket snaps into latches on
sides of instrument.
6. Make electrical connections as described in Signet 9900 Transmitter Operating
Instructions manual.
7. Close and securely latch the Accessory.
panel
Quick-clip
gasket on
front side
of panel
plugs
mounting
bracket
DC Power
Loop Voltage
3-9900.395
H COMM Module
Conduit holes
Liquid tight connectors
Latch
1. Description
2. Specifications
General
Enclosure type:
NEMA 4X
Material:
Fiberglass reinforced polyester, light gray
Hinges and Latch:
304SS
Mounting Hardware
Brackets and Screws: 304SS
Inserts:
Brass
Liquid Tight Connectors: Nylon
Approvals and Standards
• RoHS Compliant
• UL listed
• UV stabilized for outdoor use
Shown with 9900 Transmitter