Signet 8900 Multi-Parameter

SENSOR:

Select to perform calibration at the sensor via 2750 EasyCal.

INSTRUMENT: Select to perform calibration at the 8900 via the steps below.

If Cal at Instrument is selected, the following menu items appear:

Place sensor in any pH buffer. Allow suffi cient time for sensor output to stabilize, then

enter value. Maximum offset allowed: ±1.3 pH.

Place sensor in any buffer that is different by at least 2 pH units from Std value, then enter

value. Maximum offset allowed: ±1.3 pH.

pH1 Cal at:

Sensor >

Adjust the temperature measurement from the pH sensor to match a reliable reference.

Maximum offset allowed is ± 20 °C

If pH~ Cal at Instrument is selected, reset Standard, Slope and Temperature to factory

calibration.

13.2 Calibration: If Channel type = pH

ORP1 Cal at:

Sensor >

13.3 Calibration: If Channel type = ORP

Set pH1 Std:

>

Set pH1 Slope:

>

Set Temperature

T1: >

Reset pH1 to

Factory Cal: >

SENSOR:

Select to perform calibration at the sensor via 2750 EasyCal.

INSTRUMENT: Select to perform calibration at the 8900 via the steps below.

If Cal at Instrument is selected, the following menu items appear:

Place sensor in any ORP test solution. Allow suffi cient time for sensor output to

stabilize, then enter value. Maximum offset allowed: ±2900 mV.

Place sensor in any buffer that is different by at least 30 mV from Std value, then enter

value. Maximum offset allowed: ±2900 mV.

If ORP~ Cal at Instrument is selected, reset Standard and Slope to factory calibration.

Set ORP1 Std:

>

Set ORP1 Slope:

>

Reset ORP1 to

Factory Cal: >