Signet 8900 Multi-Parameter

Analog Outputs (via I/O Modules and Output Modules)

4 to 20 mA Output:
Endpoints are adjustable and reversible:
Minimum:

4.0 mA; adjustable from 3.8 to 5.0 mA

Maximum:

20.00 mA; adjustable from 19.0 to 21.0 mA

Test mode:

Produces an adjustable 4 to 20 mA signal for

functional

verifi cation of each output circuit

Isolation:

Up to 48 V AC/DC

Error condition: 22.1 mA

(default state if output source not confi gured)

Update rate:

100 mS

Accuracy: ±32

μA over entire operating temperature range

Resolution: 6

μA

Power Rejection: ±1 μA/V

Short circuit and reverse polarity protected

Passive 4 to 20 mA:
Voltage Required: 12 to 24 VDC ± 10%
Max. Impedance: 250

 @ 12 VDC

500

 @ 18 VDC

750

 @ 24 VDC

Active 4 to 20 mA:
Max Impedance: 650

0 to 5/10 VDC Output:
Output range:

0 to 5 VDC or 0 to 10 VDC, software selectable

Endpoints are adjustable and reversible:

Min. default:

0 VDC; programmable from 0 to 0.5 VDC

Max. default:

5 VDC; programmable from

4.5 to 5.5 VDC, or 9.5 to 10.5 VDC

Output load:

10 k

 minimum

Test mode:

Produces an adjustable signal for functional

verifi cation of each output circuit
Isolation:

Up to 48 V AC/DC

Error condition: 0 VDC (default state if output source not

confi gured)
Update rate:

100 mS

Accuracy:

±20 mV over operating temperature range

Resolution: 5

mV

Power Rejection: 0.5 mV/V

Environmental Requirements
Ambient operating temperature:
Back-lit LCD:

-10 °C to 55 °C (14 °F to 131 °F)

Storage Temp.:

-15 °C to 80 °C (5 °F to 176 °F)

Relative Humidity:

0 to 95%, non-condensing

Maximum Altitude:

2,000m (6,560 ft)

4,000m (13,123 ft); use only DC

power supply and, if applicable,

solid state relays to maintain UL safety

standard up to this altitude

Enclosure Rating:

NEMA 4X/IP65 (front face only)

Relay Modules
Modes of operation:
Internal Relays:

Off, Low, High, Window, Vol. Pulse,

Prop. Pulse, Pulse Width Modulation,

Total

Volume,

USP,

Advanced

External relays:

Off, Low, High, Window, USP

Total

Volume,

Advanced

Hysteresis: User

adjustable

Time Delay:

0 to 6400 seconds

Solid State Relays
Normally open/closed operation:
Software

selectable

Maximum pulse rate:

Volumetric Pulse:

600 pulses per minute

Pulse Width Modulation: 600 pulses per minute
Proportional Pulse:

400 pulses per minute

Max. voltage rating:

30 VDC or 42 VAC p-p

Current rating:

50 mA DC or 50 mA AC RMS

On-state impedance:

30 ohms or less

Off-state leakage:

400 nA or less, AC or DC

Isolation:

Up to 48 V AC/DC

Transient protection:

Embedded, up to 48 V over-voltage


Dry-contact Relays
Type:

SPDT

Form:

C

Maximum pulse rate:

Volumetric Pulse:

600 pulses per minute

Pulse Width Modulation: 600 pulses per minute
Proportional Pulse:

400 pulses per minute

Max. voltage rating:

30 VDC or 250 VAC

Current rating:

5 A

Standards and Approvals
CE

UL

Manufactured under ISO 9001 for Quality, ISO 14001 for
Environmental Management and OHSAS 18001 for
Occupational Health and Safety.
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.
China RoHS (Go to www.gfsignet.com for details)

Shipping Weights
Base Unit

1.00 kg (2.25 lb.)

Power Module

0.12 kg (0.25 lb.)

I/O Module

0.12 kg (0.25 lb.)

Output Module

0.12 kg (0.25 lb.)

Relay Module

0.12 kg (0.25 lb.)