C1 units: µs/cm – GF Signet 8900 Multi-Parameter Controller User Manual

Page 26

background image

26

Signet 8900 Multi-Parameter

Select the unit of measure for this conductivity channel:

uS/cm

mS/cm

k

• cm

M

• cm

PPM or PPB (of TDS)

If units for this channel are PPM or PPB: set the ratio of μS per PPM.

Minimum 0.01 μS per PPM

maximum 99999.9 μS per PPM

Select the unit of measure for the temperature sensor inside the conductivity electrode:

C F

Select method of temperature compensation:

None

required for USP

Linear

best for water with conductivity of more than 0.2 μS/cm (5.0 M

resistivity)

Pure Water

best for water with conductivity of less than 0.2 μS/cm (5.0 M

 resistivity)

Select decimal location for this conductivity display:

XXXXX. XXXX.X XXX.XX XX.XXX X.XXXX

Nominal setting is 2% change in conductivity for each 1ºC shift from 25ºC.

Maximum setting is 9.99% per °C

C1 Units:

µS/cm >

9.4 Channel Settings: If Channel Type = Conductivity

If this Channel is CONDUCTIVITY and the sensor is 4 to 20 mA, set the minimum and

maximum setpoints.

Dampens display, output and relay response rates for this channel.

Off: Near instantaneous updates

Low: 4 second averaging; Med: 8 second averaging; Hi: 32 second averaging.

Select decimal location for the temperature display associated with this conductivity sensor.

XXXXX. XXXX.X XXX.XX

4mA Set Point 1:

4.0000 µS >

20mA Set Point 1

20.000 µS >

C1

TDS:

2.0000 µS/PPM >

T1 Units:

C >

C1 TC Mode:

Pure Water >

C1 TC Slope:

2.00 % >

C1 Decimal:

**.*** >

T1 Decimal:

****.* >

C1 Average:

Med >