Sensor Frequency = ___________________________ Hz

Technical note:
Sensor frequency MUST be greater than 45 Hz for full scale
needle defl ection.

3. Calculate the actual max. fl ow rate in your process line as follows:

Maximum fl ow rate = Sensor frequency (step 2) × A-Factor (Section 6)

Maximum Flow Rate = __________________________________

2. Calibration

The 5090/515 fl ow metering system utilizes the AC signal
amplitude from the 515 sensor to drive the 5090 meter.

The front panel meter movement adjustment (SPAN) is easily
accessed under the removable front window for simple calibration.
A dial kit with six dial ranges, assorted fl ow unit/multiplier decals,
and dial installation instructions is included for your convenience.

1. Compatible Sensor Wiring

Frequency

counter

Red

Black

10

0

1

2

3

4

5

6

7

8

9

GPM

X100

SPAN

Signet Monitor

4. Install the appropriate dial face and fl ow unit/multiplier decal

for the maximum fl ow rate (Step 3). Refer to dial installation
instructions included with dial kit.

The dial kit includes the following:

Example: A fl ow system's maximum fl ow rate is 18.3 GPM.
The proper dial and fl ow unit/multiplier decal for this fl ow system is:
0 - 2 dial + GPM × 10 decal.

5. Disconnect frequency counter.

Access and adjust the "SPAN" potentiometer to match the
calculated maximum fl ow rate from Step 3.
After adjustment, calibration is complete.

5090

Rotor
w/magnets

Black

Red

515 Circuit

Shield

SPAN

0

1

2

3

8

6

7

5

4

9

X 100

GPM

10

SPAN

Signet Monitor

6 Dials:
0-2
0-4
0-6
0-8
0-10
0-100

Assorted Unit and Multiplier Decals:
X 10
X 100
X 1,000
X 10,000

LPM

x 1,000

GPM

x 10

m

3

/h

SPAN

0

1

2

3

8

6

7

5

4

9

10

SPAN

0

1

2

SPAN

0

1

2

SPAN

0

1

2

3

8

6

7

5

4

9

10

Signet 5090 Flow Monitor

Technical Notes:

Remove terminal block for easy wiring (Fig. 1)

Maintain cable shield through cable splice.

Route sensor cable away from AC power lines.

Use 2-conductor shielded cable for sensor cable extension up
to 61 m (200 ft).

English

Red

Blk

Shld

Sensor

Input

515

Black (AC signal out)

Red (AC signal out)

Silver (shield)

5090 Rear

Terminal

Connection

Red

Blk

Shld

Sensor

Input

Red

Blk

Shld

Sensor

Input

LISTED
77CJ

U L

Fig. 1

*3-5090.090-1*

3-5090.090-1 Rev. K 08/13

Procedure

1. Induce the maximum (stable) fl ow rate in your process line.

2. Using the frequency counter, monitor and record the 515

sensor frequency output:

Equipment Required
• Frequency

counter

Signet 515 Sensor installed in process line

Standard and phillips head screwdriver