GF Signet 5090 Sensor-Powered ProPoint Flow Monitor User Manual
Signet 5090 flow monitor, Calibration, Compatible sensor wiring
Sensor Frequency = ___________________________ Hz
Technical note:
Sensor frequency MUST be greater than 45 Hz for full scale
needle defl ection.
3. Calculate the actual max. fl ow rate in your process line as follows:
Maximum fl ow rate = Sensor frequency (step 2) × A-Factor (Section 6)
Maximum Flow Rate = __________________________________
2. Calibration
The 5090/515 fl ow metering system utilizes the AC signal
amplitude from the 515 sensor to drive the 5090 meter.
The front panel meter movement adjustment (SPAN) is easily
accessed under the removable front window for simple calibration.
A dial kit with six dial ranges, assorted fl ow unit/multiplier decals,
and dial installation instructions is included for your convenience.
1. Compatible Sensor Wiring
Frequency
counter
Red
Black
10
0
1
2
3
4
5
6
7
8
9
GPM
X100
SPAN
Signet Monitor
4. Install the appropriate dial face and fl ow unit/multiplier decal
for the maximum fl ow rate (Step 3). Refer to dial installation
instructions included with dial kit.
The dial kit includes the following:
Example: A fl ow system's maximum fl ow rate is 18.3 GPM.
The proper dial and fl ow unit/multiplier decal for this fl ow system is:
0 - 2 dial + GPM × 10 decal.
5. Disconnect frequency counter.
Access and adjust the "SPAN" potentiometer to match the
calculated maximum fl ow rate from Step 3.
After adjustment, calibration is complete.
5090
Rotor
w/magnets
Black
Red
515 Circuit
Shield
SPAN
0
1
2
3
8
6
7
5
4
9
X 100
GPM
10
SPAN
Signet Monitor
6 Dials:
0-2
0-4
0-6
0-8
0-10
0-100
Assorted Unit and Multiplier Decals:
X 10
X 100
X 1,000
X 10,000
LPM
x 1,000
GPM
x 10
m
3
/h
SPAN
0
1
2
3
8
6
7
5
4
9
10
SPAN
0
1
2
SPAN
0
1
2
SPAN
0
1
2
3
8
6
7
5
4
9
10
Signet 5090 Flow Monitor
Technical Notes:
•
Remove terminal block for easy wiring (Fig. 1)
•
Maintain cable shield through cable splice.
•
Route sensor cable away from AC power lines.
•
Use 2-conductor shielded cable for sensor cable extension up
to 61 m (200 ft).
English
Red
Blk
Shld
Sensor
Input
515
Black (AC signal out)
Red (AC signal out)
Silver (shield)
5090 Rear
Terminal
Connection
Red
Blk
Shld
Sensor
Input
Red
Blk
Shld
Sensor
Input
LISTED
77CJ
U L
Fig. 1
*3-5090.090-1*
3-5090.090-1 Rev. K 08/13
Procedure
1. Induce the maximum (stable) fl ow rate in your process line.
2. Using the frequency counter, monitor and record the 515
sensor frequency output:
Equipment Required
• Frequency
counter
•
Signet 515 Sensor installed in process line
•
Standard and phillips head screwdriver