GF Signet 2721 Preamplifier User Manual

Signet 2721 ph/orp preamplifi er, English

background image

Signet 2721 pH/ORP Preamplifi er

305 mm/

12 in.

30 mm/

1.2 in.

43 mm/

1.7 in.

BNC

connector

1. Specifi cations

Housing material:

Low density polyethylene

Electronics assembly:

Epoxy encapsulated

Input Impedance:

> 10

11

Ω

Operating temp.:

-15 to 65 °C

Gain:

X1

(unity)

Input power:

± 4.5 to ± 8 VDC, dual supply

Current consumption:

< 1 mA, dual supply

mV input signal:

< ± 2500 mV

2. pH

Wiring

3-2721.090

Rev. F 10/12

English

*3-2721.090*

English

• 8750 terminal numbers vary depending on model number. See 8750 Instruction Manual.
• A 3 kΩ resistor (not supplied) across the T+ and T- terminals of the Signet pH instrument provides a simulated temperature input that

is required for pH operation.

• A 10 kΩ resistor (not supplied) across the T+ and T- terminals of the Signet instrument initiates ORP operation.
• A shielded cable/preamplifi er enclosure is recommended for optimum performance.

Dimensions:

3 kΩ

V-

T-

10 kΩ

Signet 8750

Signet 8750

V-
V+
mV Input

Iso. Gnd

T+
T-

V+
mV Input

Iso. Gnd

T+

Black
Red
Brown
Green

Black
Red
Brown
Green

3. ORP

Wiring

Use shielded cable for cable extensions up to 120 m (400 ft.)

Black

Red

Brown

Green

Iso. Gnd

4. Cable

Extensions