Signet 9900 Direct Conductivity/Resistivity Module Instruction Sheet

3-9900.092 Rev. A 4/11 English

● Terminals accept 16 to 28 AWG wire.

● Sensor cable length can be extended to 30 m (100 ft) (except 2819

and 2839, limited to 4.6 m (15 ft).

● Strip 10 to 12 mm (0.4 to 0.5 in.) of insulation from wire tips

and tin bare ends to eliminate fraying.

● Insert wire tip or ferrule completely into the terminal and

secure with the screw.

● Do not allow any AC leads that may be connected to the

internal relays to come in contact with low voltage wiring.

Georg Fischer Signet LLC, 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: gfsignet.com • Or call (in the U.S.): (800) 854-4090
For the most up-to date information, please refer to our website at www.gfsignet.com

3-9900.092 Rev. A 04/11 English

© Georg Fischer Signet LLC 2011

*3-9900.092*

Field Mount and Panel Mount
are identical.

2819-2823
2839-2842

If the 9900 Base Unit will be mounted in a panel, plug-in modules may be
installed either before or after the base unit is mounted. If the 9900 Base
Unit will be mounted using the accessory wall mount bracket, install plug-in
modules fi rst.
The Conductivity/Resistivity module plugs in and attaches with a screw.

1. It may be helpful to use a standard screwdriver to gently pry

removable terminals away from the receptacle on the plug-in modules.
Reverse this step to reinstall the panel after the plug-in modules are
installed.

2. Remove terminals from the receptacle of each plug-in module prior to

installation.

To interface a stand-alone conductivity sensor (2819-2823 or 2839-2842) to the 9900, the 3-9900.394 Direct
Conductivity Module must be used.

CAUTION:

Exercise care when

installing module.
Do not bend connecting pins.

Align pins and connectors then push
module fi rmly into place, then attach with
screw.
Avoid Electrostatic Discharge (ESD):

● Minimize handling of module to

reduce the possibility of damage
due to ESD.

● Handle module by the edges. Never

touch any exposed circuitry
or contacts.

● Wear an anti-static wristband or

stand on an anti-static mat, or
keep one hand touching a properly
grounded pipe or other properly
grounded piece of metal when
handling module.

Mfr. Part No. Code

Description

3-9900.394

159 001 669

Direct Conductivity/Resistivity Module

6682-0204

159 001 709

Conductivity Module Plug, 4 Pos, Right Angle