Signet 2850 Conductivity/Resistivity Sensor Electronics

English

1. Description

• Signet 2850 Conductivity/Resistivity Sensor Electronics provides either a two-wire 4 to 20 mA output or Digital (S

3

L) format.

• The 4 to 20 mA output models provide eight ranges for each electrode cell constant, plus the ability to invert each range.
• The EasyCal feature allows the devices to automatically recognize standard conductivity test solution values for simple fi eld calibration.
• The Conductivity sensor provided with integral systems will have its custom cell constant information programmed into the electronics

at the factory to provide a 1% sensor accuracy. See Section 8 for details.

SAFETY PRECAUTIONS
1. Depressurize and vent system prior to installation or removal.
2. Confi rm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifi cations.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents, care should be taken and appropriate

eye, face, hand, body, and/or respiratory protection should be used.

2. Specifi cations

General
Compatible Electrodes:

Any Signet Conductivity/
Resisitivity electrode

(2818-2823,

2839-2842)

Temperature Compensation: PT-1000 RTD (2% per °C)
Pure Water Compensation:

Auto-switching when using
0.01 cell and raw conductivity
value < 0.5 μS

EasyCal:
Automatic recognition of the following conductivity values:
• 146.93 μS, 1408.8 μS, 12856 μS (@ 25 °C)
(Test solutions per ASTM D1125-95)
• 10 μS, 100 μS, 200 μS, 500 μS, 1000 μS, 5000 μS, 10,000 μS,

50,000 μS, 100,000 μS (@ 25 °C)

Electrical
Power:

12 to 24 VDC (4 to 20 mA output)

5 VDC ±5% regulated, 3.0 mA
max. for Digital (S

3

L) output

(Reverse polarity and short circuit
protected)

Digital (S

3

L) Output:

Serial ASCII, TTL level 9600 bps

• Accuracy:

Conductivity: ±2% of reading

Temperature: ±0.5 °C

• Resolution:

Conductivity: 0.1% of reading

Temperature: < 0.2 °C

• Update Rate:

Single input models: < 600 ms

Dual input models: < 1.2 s

Current Output:
• Field-selectable ranges
• Factory set span:
0.01 cell (2818, 2819, 2839): 4 to 20 mA = 0 to 100 μS
0.10 cell (2820, 2840):

4 to 20 mA = 0 to 1000 μS

1.0 cell (2821, 2841):

4 to 20 mA = 0 to 10,000 μS

10.0 cell (2822, 2842):

4 to 20 mA = 0 to 200,000 μS

20.0 cell (2823):

4 to 20 mA = 0 to 400,000 μS

• Max. Loop Resistance:

50 Ω @ 12 VDC

325

Ω @ 18 VDC

600

Ω @ 24 VDC

• Accuracy:

±2% of output span

• Resolution:

7 μA

• Update Rate:

<600 ms

• Error Indication:

22 mA

Available data via Digital (S

3

L) Output:

• Raw conductivity
• Calibrated conductivity
• Calibrated temperature-compensated conductivity
• Temp. Error Indication:

Open input and out-of-range
diagnostics for temperature or
internal electronic error.

Environmental
Material: PBT
Enclosure Rating:

NEMA 4X/IP65

Operating Temperature:

-10 to 85 °C (14 to 185 °F)

Storage Temperature:

-20 to 85 °C (-4 to 185 °F)

Relative Humidity:

0 to 95%, non-condensing

Shipping Weight:

0.75 kg (1.75 lb.)

Standards and Approvals
• CE
• Manufactured under ISO 9001 and ISO 14001

3-2850.090-1 Rev.

H

06/12 English

*3-2850.090-1*

Contents

1. Description
2. Specifi cations
3. In-Line

Installation

4. Submersible

Installation

5. Wiring 4 to 20 mA Output version
6. Wiring Digital (S

3

L) Output version

7. Wiring Dual Input Digital (S

3

L) Output version

8. Cell Constant Selection
9. Range Selection for 4 to 20 mA Output
10. Calibration
11. Maintenance
12. Ordering Information