Dsp menu, Part 3 - keypad configuration, Display menu – Dynasonics TFXP Series Transit Time Ultrasonic Flow Meter User Manual

Page 77

background image


Rev. 5/09

-3.39-

TFXP

Correction
Factor

Graphics
Display Mode

the meter. This is not as desirable as the Set Zero and should only
be used with large pipes.

1. Press EDIT at the Reset Zero prompt, use the arrow keys to

make the display read Yes.

2. Press ACCEPT.

COR FTR -- Correction Factor (Value)

(0.500 – 1.500)


This function can be used to make the TFX system agree with a
different or reference flow meter by applying a correction factor/
multiplier to the readings and outputs. A factory calibrated system
should be set to 1.000 (no correction factor). The range of settings
for this entry is 0.500 to 1.500. The following example describes a
possible use for the Correction Factor entry.

Example: The TFX meter is indicating a flow rate that is 4% higher
than another flow meter located in the same pipe line. To make the
TFX indicate the same flow rate as the other meter, enter a
Correction Factor of 0.960 to lower the readings by 4%.

DISPLAY LINES -- Number of Display Lines

1. 2 Lines
2. 4 Lines

Allows the selection of a two line or four line display format on the
graphics display module.

In 2 line mode, the display will show flow measurements with larger
characters on the top half of the window and smaller standard sized
characters on the lower half of the window. In 4 line mode, the
display will show flow measurements with standard sized characters
on four lines in the window.

8.

DSP MENU

--

DISPLAY MENU

PART 3 - KEYPAD CONFIGURATION