PART 1 - SPECIFICATIONS

TRANSMITTER

Power Requirements

Internal 12 V lead-acid Gel Cell battery provides 24 hrs of continuous operation @ 20 °C. Charging: Wall mount
power converter. 115 or 230 VAC 50/60 Hz ±15% VA max; 12-15 VDC @ 2.5 VA max

Velocity Range

-40 to +40 FPS (-12 to +12 MPS)

Inputs/Outputs

All output modules are optically isolated from earth and system grounds.
One module and one data logger may be installed.

Standard Options

4-20 mA

800 Ohm max; 12-bit resolution, passive or active

Data Logger

200k events in 30k pages, 16-bit, integral DB-9 RS232C connection, can be removed and installed without
disconnecting system power, data transfer rates to 57.6k

Rate Pulse

MOSFET, 0.21 Ohms, 100 V max, 0 to 2,500 Hz max

Dual Relay

2 separate Form C relays, 200 VAC max at 0.5 A (resistive)

RS232C

Data rate to 57.6k

RS485

Supports up to 119 drops

Heat Flow (RTD)

Supports two 1000 Ohm RTDs, multiplexed, 12-bit resolution

Display

128 x 64 pixel graphics LCD, LED backlit. Two user selectable font sizes 0.35" (8.9 mm) or 0.2" (5 mm)
8 digit rate, 8 digit totalizer (resettable)

Units User

configured

Rate

Gal, liters, million gal, ft³, m³, acre-ft, oil barrels (42 gal), liquid barrels (31.5 gal), ft, m, lb, kg, BTU, MBTU,
MMBTU, ton

Time

Seconds, minutes, hours, days

Totalizer

Gal, liters, million gal, ft³, m³, acre-ft, oil barrels (42 gal), liquid barrels (31.5 gal), lb, kg, BTU, MBTU, MMBTU, ton

Mode

Forward, reverse, net, batch

Ambient Conditions

-40 °F to +185 °F (-40 °C to +85 °C), 0 to 95% relative humidity (non-condensing)

Enclosure

NEMA 4X (IP-66) while open, NEMA 6 (IP-68) while closed, ABS with SS hardware

Size

14.00" H × 6.06" W × 10.56" D (355.6 mm H × 153.9 mm W × 268.2 mm D); 14.8 lbs (6.7 Kg)

Flow Rate Accuracy

DTTN / DTTH: ±1% of reading at rates > 1 FPS (0.3 MPS), ±0.01 FPS (±0.003 MPS) at rates lower than 1 FPS
DTTS/DTTC: 1" and larger units, ±1% of reading from 10-100% of measuring range, ±0.01 FPS (±0.003 MPS)
at rates lower than 10% of measuring range; 1/2" and 3/4" units, ±1% FS

Flow Sensitivity

0.001 FPS (0.0003 MPS)

Repeatability ±0.01%

of

reading

Response Time (Flow) 0.3

to

30 seconds, user configured, to 100% of value, step change in flow

Security

Keypad lockout, user selected 4 digit access code

Approvals

Ordinary areas

Software

ULTRALINK™, compatible with Window

®

98/2000/XP/Vista

®

Liquid Types

Most non-aerated, clean liquids

Cable Length

Up to 990 ft (300 meters); standard lengths 20, 50, 100 ft (6, 15, 30 meters)

Pipe Sizes

DTTN / DTTH: 2 inch and larger
DTTS / DTTC (small pipe): 1/2", 3/4", 1", 1-1/4", 1-1/2" (ANSI pipe, copper tube, tube)

Environment

NEMA 6

Pipe Surface
Temperature

DTTN / DTTC: -40 °F to +250 °F (-40 °C to +121 °C)
DTTS: -40 °F to +185 °F (-40 °C to +85 °C)
DTTH: -40 °F to +350 °F (-40 °C to +177 °C)

Ambient Conditions

-40 °F to +185 °F (-40 °C to +85 °C)

Housing Material

DTTN / DTTC: CPVC, Ultem

®

, and nylon

DTTS: PVC, Ultem

®

, and nylon

DTTH: PTFE, Vespel

®

, and nickel-plated brass

Approvals

Standard: None
Optional - DTTN only: CSA Class I, Div 1, Groups C & D; requires intrinsically safe transducer kit with barrier

TRANSDUCERS

Other Options