Part 2 - transducer & rtd installation – Dynasonics TFXP Series Transit Time Ultrasonic Flow Meter User Manual

Page 29

background image


Rev. 5/09

- 2.11 -

TFXP

4. If after adjustment of the transducers the signal strength does

not rise to above 5 percent, then an alternate transducer
mounting method should be selected. If the mounting method
was W-Mount, then reconfigure the TFX for V-Mount, reset the
TFX, move the downstream transducer to the new location and
repeat Step 3 on page 2.10.

5. Certain pipe and liquid characteristics may cause signal strength

to rise to greater than 195 percent. The problem with operating a
TFX with very high signal strength is that the signals may
saturate the input amplifiers and cause erratic readings. To
decrease the signal strength, move one transducer a small
distance radially around the pipe, as shown in Figure 2.6.

Figure 2.6 — High Signal Strength Correction

PART 2 - TRANSDUCER & RTD INSTALLATION