Rev. 5/09

-3.38-

TFXP

Setting/
Calibrating
Zero Flow

SET ZERO -- Calibrating Zero Flow (Choice)

1. No

2. Yes


Because every flow meter installation is slightly different and sound
waves can travel in slightly different ways through these various
installations, it is important to remove the zero offset at zero flow to
maintain the meter’s accuracy. A provision is made using this entry
to establish “Zero” flow and eliminate the offset.

To zero the meter:

1. The pipe must be full of liquid.

2. Flow must be absolute zero – securely close any valves and

allow time for any settling to occur.

3. Press EDIT at the Set Zero prompt, use the arrow keys to make

the display read Yes.

4. Press ACCEPT.

The procedure is complete.

D-FLT 0 -- Setting a Zero Flow Point (Choice)

1. No
2. Yes


If the flow in a piping system cannot be shut off, allow the Set Zero
procedure described above to be performed, then the default zero
should be utilized. This procedure places an actual value of zero in

MINIMUM

RATE SETTING

MAXIMUM

RATE SETTING

SUBSTITUTE

FLOW SETTING

DISPLAY

READING

DURING ERRORS

0.0

1,000.0

0.0

0.000

-500.0

500.00

50.0

0.000

-100.0

200.0

33.3

0.000

0.0

1,000.0

-5.0*

-50.00

* NOTE: ULTRALINK ™ is required to set values outside of 0.0 - 100.0.

Table 3.8 — Typical Zero Settings

Factory
Default Zero
Calibration