The SEN MENU is utilized to select the type of transducer that will
be interfaced with the TFX. Select the appropriate transducer from
the list and save the configuration. If the transducer selection is
modified, a system reset is required.

XDUCER TYP -- Transducer Type (Choices and Values)

1. DTTN Standard Transducers (DTTN

Clamp-On)

2. DTTH High Temperature Transducers (DTTH

Clamp-On)

3. DTT1500

(DTT

1500)

4. Small Pipe Copper

(DTTSnC)

5. Small Pipe ANSI

(DTTSnP)

6. Small Pipe Tubing

(DTTSnT)

The SEC MENU allows the user to make password revisions, reset
the flow totalizer and reset the transmitter microprocessor.

TOT RES -- Totalizer Reset (Choice)

1. NO
2. YES


Select YES to reset all flow totalizers/accumulators to zero.

SYS RSET -- System Reset (Choice)

1. NO
2. YES

Select YES to initiate a microprocessor reset.

NOTE: All system configurations and totalizer values will be
retained.



5.

SEN MENU

--

SENSOR MENU

6.

SEC MENU

--

SECURITY MENU