Part 3 - keypad configuration, Engineering units total exponent minimum flow rate – Dynasonics TFXP Series Transit Time Ultrasonic Flow Meter User Manual

Page 53

background image


Rev. 5/09

-3.15-

TFXP

Engineering
Units
TOTAL
Exponent

Minimum
Flow Rate

13. Millions of BTUs (MMBTU)
14.
Tons

(TON)


Select a desired engineering unit for flow accumulator (totalizer)
measurements.

TOTL E -- Flow Totalizer Exponent Value (Choice)

E-1 to E6


Utilized for setting the flow totalizer exponent. This feature is useful
for accommodating a very large accumulated flow. The exponent is
a ×10

n

multiplier, where "n" can be from -1 (×0.1) to +6

(×1,000,000). Table 3.5 should be referenced for valid entries and
their influence on the TFX display.

MIN RATE -- Minimum Flow Rate Settings (Value)

Rate Unit/Rate Interval


A minimum volumetric flow rate setting is entered to establish filter
software settings. Volumetric entries will be in the Engineering Rate
Units and Interval selected on pages 3.13 and 3.14. For
unidirectional measurements, set Min Rate to zero. For bi-
directional measurements, set Min Rate to the highest negative
(reverse) flow rate expected in the piping system.

Exponent

Display Multiplier

E-1

× 0.1

E0

× 1 (no multiplier)

E1

× 10

E2

× 100

E3

× 1,000

E4

× 10,000

E5

× 100,000

E6

× 1,000,000

Table 3.5 — Totalizer Exponent Values

PART 3 - KEYPAD CONFIGURATION