Part 3 - keypad configuration – Dynasonics TFXP Series Transit Time Ultrasonic Flow Meter User Manual
Page 47

Rev. 5/09
-3.9-
TFXP
Liner
Sound Speed
LINER SS -- Speed of Sound in the Liner (Value)
1. ENGLSH
(Feet per Second)
2. METRIC
(Meters per Second)
Allows adjustments to be made to the speed of sound in the liner. If
the Units value was set to Englsh, the entry is in FPS (feet per
second). Metric entries are made in MPS (meters per second).
If a liner was chosen from the Liner Mat list, a nominal value for
speed of sound in that media will be automatically loaded. If the
actual sound speed rate is known for the pipe liner and that value
varies from the automatically loaded value, the value can be
revised.
FL TYPE -- Fluid/Media Type (Choice)
1. Water Tap
(WATER)
2. Sewage (SEWAGE)
3. Acetone (ACETONE)
4. Alcohol (ALCOHOL)
5. Ammonia
(AMMONIA)
6. Benzene
(BENZENE)
7. Brine
(BRINE)
8. Ethanol (EHTANOL)
9. Ethylene Glycol (ETH-GLYC)
10. Gasoline
(GASOLINE)
11. Glycerin (GLYCERIN)
12. Isopropyl Alcohol (ISO-ALC)
13. Kerosene
(KEROSENE)
14. Methanol
(METHANOL)
15. Oil Diesel
(DIESEL)
16. Oil Hydraulic
(HYD OIL)
[Petro-based]
17. Oil Lubricating (LUBE
OIL)
18. Oil Motor
(MTR
OIL)
[SAE 20/30]
19. Water Distilled (WATR-DST)
20. Water Sea
(WATR-SEA)
21. Other
(OTHER)
This list is provided as an example. Additional fluid types may have
been added. Select the appropriate liquid from the list or select
Other if the liquid is not listed.
Fluid
Type
PART 3 - KEYPAD CONFIGURATION